Entry   Pressure On mirror Chambers Plot, posted by Manar Amer at ThomX igloo about vacuum PressureOverTime.png
    Reply   Pressure On mirror Chambers Plot, posted by Manar Amer at ThomX igloo about vacuum 
       Reply   Pressure On mirror Chambers Plot, posted by Manar Amer at ThomX igloo about vacuum 
Message ID: 31     Entry time: Fri Oct 22 17:36:05 2021     Reply to this: 32
 Author: Manar Amer 
 Status: Fixed 
 Type: info 
 Category: vacuum 
 Location: ThomX igloo 
 Title: Pressure On mirror Chambers Plot 

A plot showing the descend of the vacuum in the mirror chambers (data to plot it obtained yesterday evening)

The regions where there is a drastic decrease is when the ionic pump was turned on.

Today in the morning the pressure was

  • on the left                    :  1.3*10^-9 mbar
  • on the right (sortie X)  :  4.9*10^-9 mbar

in the evening the pressure was

  • on the left                    :  1.4*10^-9 mbar
  • on the right (sortie X)  :  4.7*10^-9 mbar

Attachment 1: PressureOverTime.png  46 kB  | Hide | Hide all