Entry   Oscilloscopes control room panel, posted by Ronic Chiche at Optical room about software 

today with Philippe Gauron and Olivier Neuveu, we tested the control room TANGO panel for the oscilloscopes.
the ThomX IP addresses of the scopes will be : 192.168.229.27 and 192.168.229.28

this panel allow ONLY to read a measurement of the amplitude of the channels 1, 2 and 3 (laser power, cavity transmission and reflection) of the oscilloscope.
this is just a monitoring panel.

at the TANGO panel start, each channel of the scope is initialized : DC coupling, 1Mohms, 100mV/div, 20MHz BW, -4 Div position, 0V offset
and the acquisition setup is : 100 ms/div, 1000 points
and an amplitude measurement is done for each channel

if one needs to remotely control the scopes, one can install NI Labview SignalExpress TektroniX Edition which allow a full remote control/reading of the scope from a local PC.
use the menu Add Step -> Tektronix -> Acquire Signals -> Tek DPO/MSO 4000  to lauch the scope panel in SignalExpress.

one has to stop the TANGO server of the scope panel before using SignalExpress
and vice-versa, one has to stop SignalExpress running and untick the "SignalExpress controls instrument" to let the TANGO panel access the scope.