today with Philippe Gauron and Olivier Neuveu, we tested the control room TANGO panel for the oscilloscopes.
the ThomX IP addresses of the scopes will be : 192.168.229.27 and 192.168.229.28
this panel allow ONLY to read a measurement of the amplitude of the channels 1, 2 and 3 (laser power, cavity transmission and reflection) of the oscilloscope.
this is just a monitoring panel.
at the TANGO panel start, each channel of the scope is initialized : DC coupling, 1Mohms, 100mV/div, 20MHz BW, -4 Div position, 0V offset
and the acquisition setup is : 100 ms/div, 1000 points
and an amplitude measurement is done for each channel
if one needs to remotely control the scopes, one can install NI Labview SignalExpress TektroniX Edition which allow a full remote control/reading of the scope from a local PC.
use the menu Add Step -> Tektronix -> Acquire Signals -> Tek DPO/MSO 4000 to lauch the scope panel in SignalExpress.
one has to stop the TANGO server of the scope panel before using SignalExpress
and vice-versa, one has to stop SignalExpress running and untick the "SignalExpress controls instrument" to let the TANGO panel access the scope. |