32
|
Mon Apr 8 11:55:40 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | P4 cleaned and finesse measurement |
P4 cleaned with isopropanol.
Finesse measured afterward: ~=13 000 (clear increase) |
|
31
|
Mon Apr 8 10:18:31 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Finesse after 2days |
Finesse measured this morning after P1 SUPRASIL being into the cavity for two days. In order to check if it's getting dirty.
Finesse still ~= 11 000 |
|
30
|
Fri Apr 5 11:28:39 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Finesse with P1 SUPRASIL |
P1 SUPRASIL has been installed instead of P1 Sapphire. It has been cleaned 3 times on each face with wipe+isopropanol before to be installed.
Finesse ~= 10 000 - 11 000 (EOM Sweep of 100 kHz in 10sec, see last image.)
Coupling ~= 45%
Motors position:
P1 = -825 000
S2 = +825 000
S3 = +825 000
P4 = -825 000 |
| Attachment 1: 190404_mode_size_with_P1_SUPRASIL.PNG
|
 |
| Attachment 2: 190404_mode_fit_with_P1_SUPRASIL.PNG
|
 |
| Attachment 3: tek00000.png
|
 |
| Attachment 4: tek00002.png
|
 |
|
29
|
Fri Apr 5 11:18:47 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Transmission of the mirrors |
P1 transmission out of the mount (means angle of the mirror is not the real angle):
Pin = 3.2 mW
Pout = 545 nW
Offset = 68 nW
Transmission = (545-68)/3.2e6 = 149 ppm
| Loïc Amoudry wrote: |
|
After verification, transmission of P1 correspond to ~150ppm.
The 1337ppm come from a beam leakage. Reflection from P1 goes to the input window which reflects again a part of the beam directly into the cavity between P1 and S3 then between P4 and S2.
| Loïc Amoudry wrote: |
|
We openned the cavity, removed S2 and measured trans of P1 and same for S2.
Pin ~= 3.91 mW
PS2 ~= 20 nW --> ~~ 5 ppm !!!
PP1 ~= 5.23 uW --> ~~ 1337 ppm !?!?!
P1 is the Sapphire mirror. 3mm thick. Tomorrow we'll measure its trans again, being out of the mount for constraint issues.
|
|
|
|
28
|
Thu Apr 4 09:42:12 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Transmission of the mirrors |
After verification, transmission of P1 correspond to ~150ppm.
The 1337ppm come from a beam leakage. Reflection from P1 goes to the input window which reflects again a part of the beam directly into the cavity between P1 and S3 then between P4 and S2.
| Loïc Amoudry wrote: |
|
We openned the cavity, removed S2 and measured trans of P1 and same for S2.
Pin ~= 3.91 mW
PS2 ~= 20 nW --> ~~ 5 ppm !!!
PP1 ~= 5.23 uW --> ~~ 1337 ppm !?!?!
P1 is the Sapphire mirror. 3mm thick. Tomorrow we'll measure its trans again, being out of the mount for constraint issues.
|
|
|
27
|
Tue Apr 2 18:05:46 2019 |
Loïc Amoudry | Fixed | info | lasers and optics | Optical room | Beam size at closest spherical mirrors position |
S3 and S2 @ +1 640 000 and P1 and P4 @ -825 000
Diameter in x: 4.7mm
Diameter in y: 4.7mm |
|
26
|
Tue Apr 2 18:03:31 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Transmission of the mirrors |
We openned the cavity, removed S2 and measured trans of P1 and same for S2.
Pin ~= 3.91 mW
PS2 ~= 20 nW --> ~~ 5 ppm !!!
PP1 ~= 5.23 uW --> ~~ 1337 ppm !?!?!
P1 is the Sapphire mirror. 3mm thick. Tomorrow we'll measure its trans again, being out of the mount for constraint issues. |
|
25
|
Tue Apr 2 17:59:18 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Finesse after removing wires |
19/04/01: We removed the wires from the beam path in the cavity. Finesse stayed at 3400. FSR: 33.342 MHz.
When locked:
Input Power: ~4mW
Through S2: ~0.950uW
Through S3: ~0.930uW
Through P4: ~0.850uW |
|
24
|
Fri Mar 29 18:33:15 2019 |
Ronic Chiche | Fixed | issue | mechanics | lasers and optics | Optical room | Alignment procedure |
we did again the complete alignment procedure starting with iris and optimizing injection motors Ma and Mb, then installing mirror S2, S3, P4 and P1, optimizing thetaX and thetaY axis for each motor.
everything was fine until we installed P1. we tried to optimize thetaX and thetaY of P1 and we clearly observed a strange motion when doing that :
- for thetaX axis, steps in one direction seem to have a different length from steps in reverse direction.
- for thetaY axis, trying to move in one direction, makes sometime a motion in the reverse direction.
at this moment, it is difficult to say if the problem comes from the controller, the mirror mount or the motor itself.
if both axis are concerned on the same mirror (P1), maybe the problem comes from the mount... to be continued on Monday.
|
|
23
|
Thu Mar 28 09:25:58 2019 |
Loïc Amoudry | Fixed | issue | mechanics | Optical room | Motors Z axis reversed |
All Z motors (for P1, S2, S3 and P4) have been initialized again with a proper HOME command with their own controller with the "superviser" software (which means they all moved to the +1650000 position).
Thus, they have been moved to their default position : +825000 for spherical mirrors (S2 and S3) and -825000 for plan mirrors (P1 and P4).
reminder : lower values on the motors makes the motor tip going inside which makes the cavity length larger.
thus spherical mirrors are in default position with a possible big range to put inside the motor tip (for thermal effects compensation).
by complementarity, plan mirrors are in default position with a possible big range to put outside the motor tip.
the Labwindows CVI was buggy about the proper representation of the motors motion in Z direction (wrong direction and wrong default position).
the software is now fixed.
| Loïc Amoudry wrote: |
|
S2 and S3 z-axis are initially at 1/4 (5mm/20mm) of there range over ~20mm range. This is meant to be able to spread them up to ~15mm in case of thermal effect.
Actually, they are in an opposite position. Means that they can only be spread of 5mm from the initial position.
|
|
|
22
|
Tue Mar 26 18:32:40 2019 |
Loïc Amoudry | Fixed | report | mechanics | lasers and optics | cabling | Optical room | Wires in laser path |
There was some ceramic wires close to the beem path right before P4. We moved them away (we can imagine they could move again with vibration). |
| Attachment 1: before.jpg
|
 |
| Attachment 2: after_wire_displaced.jpg
|
 |
|
21
|
Tue Mar 26 18:26:32 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Check of mirrors and coating directions |
Direction of HR face of P1 and P4 have been checked, there was no problem. The mirrors are in the good direction and the coating is on the good face.
Once P1 and P4 outside of the cavity, we checked the coating position by sending the HeNe laser with an angle. There was two reflections and the most intense was the one closer to the normal to the mirror.
On P4 we noticed a spot which seems to be in the coating (does not move while sending nitrogen). |
| Attachment 1: ThomX_P4.jpg
|
 |
|
20
|
Fri Mar 22 17:53:10 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | vacuum | Optical room | Cavity at atm pressure |
Dry air injected into the cavity.
Lock has been proceeded without any trouble. ~~4000 finesse. Considered to be the same than before.
x and y motors axis put back to the last "cavity at air" position. z position kept at maximum distance between spherical mirrors. |
| Attachment 1: finessealair.isf
|
|
19
|
Tue Mar 19 17:01:52 2019 |
Loïc Amoudry | Fixed | issue | mechanics | Optical room | Motors Z axis reversed |
S2 and S3 z-axis are initially at 1/4 (5mm/20mm) of there range over ~20mm range. This is meant to be able to spread them up to ~15mm in case of thermal effect.
Actually, they are in an opposite position. Means that they can only be spread of 5mm from the initial position. |
|
18
|
Tue Mar 19 16:35:32 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Beam size and finesse |
FSR initiale : 33.34 MHz (S3 = -825000 et S4 = -825000)
Beam size initiale : wx=2.05mm et wy=2.15mm
Finesse initiale : 3400
FSR après éloignement des miroirs sphériques : 33.29 MHz (S3 ~= -1496000 et S4 ~= -1496000)
Beam size : wx=1.7mm et wy=1.85mm
Finesse : 3600
FSR après rapprochement des miroirs sphériques : 33.39 MHz (S3 ~= -250000 et S4 ~= -500000)
Beam size : wx = 2.1mm et wy=2.4mm
Fit de la caméra jamais au dessus de 90%. ~~80%. |
|
17
|
Tue Feb 19 15:50:08 2019 |
Loïc Amoudry | Fixed | info | mechanics | Optical room | Motors connexions |
Motors are connected. The true mirror's name are P1 for injection mirrors, S2, S3 and P4. (S=spherical, P=planar).
They are still some old schemes (and notations on the cavity) which mention M1, M2, M3 and M4. I make the link here:
M3 = P1
M1 = S2
M2 = S3
M4 = P4
The wires are plugged as follows (I use notations wrote on the cavity and wires):
M3x = 1-1
M3y = 1-2
M3z = 2-1
M1x = 1-3
M1y = 1-4
M1z = 2-2
M2x = 1-5
M2y = 1-6
M2z = 2-3
M4x = 1-7
M4y = 1-8
M4z = 2-4 |
| Attachment 1: Validation_moteurs.jpg
|
 |
|
16
|
Thu Feb 14 18:20:52 2019 |
Loïc Amoudry | Fixed | report | mechanics | lasers and optics | Optical room | Basic alignment & alignment motors check |
14/02/19: The 4 motors but Mby control works. Mby motor home does not bring it to its "middle" position but to the end of its range. Issue circumvented using the higher rights software.
The beam is now going through the cavity windows (from P1 to S2). |
|
15
|
Fri Feb 1 14:10:25 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Other | Mirrors sent to LMA |
30/01/19 - As LMA people came, we discussed about the spots on ThomX mirrors which should be clean. As mentionned in "SBOX commissioning", these spots are normal.
An other question appeared during the conversation about the storage time and location of these mirrors. They have been stored for more than two years in plastic boxes. LMA people explained that some company do not let anything more than 6 months in these boxes for pollution issues.
We merged into the same conclusion, clean the mirrors in case of pollution. Finally, they took the mirrors and we will go to LMA to recover them between 11/02 and 14/02. |
|
14
|
Fri Feb 1 13:39:21 2019 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Mirrors checked with binocular microscope |
29/01/19 - ThomX mirrors have also been checked with binocular microscope. They shown the same spots as S-BOX mirrors.
1st image shows reflective face of a spherical mirror.
2nd image shows reflective face of M1 (sapphire). |
| Attachment 1: 41_THOMX_SphericalZoom1x.jpg
|
 |
| Attachment 2: 49_THOMX_M1ReflectiveZoom4x.jpg
|
 |
|
13
|
Fri Feb 1 13:20:47 2019 |
Loïc Amoudry | Fixed | report | utilities | Optical room | End of cleaning, plastic fixed on cavity |
08/02/19 - End of the cleaning process. |
| Attachment 1: covered_by_plastic.png
|
 |
| Attachment 2: covered_by_plastic2.png
|
 |