  459   Tue Oct 28 10:07:48 2025 Ronic ChicheFixedissuelasers and optics | detectors and electronicsThomX iglooHigh frequency noiseAfter this observation, we did a test with
Daniele to try to correlate the road traffic
in front of the Igloo with the lock losses
  
  460   Tue Oct 28 17:31:47 2025 Ronic ChicheFixedissuelasers and optics | detectors and electronicsThomX iglooHigh frequency noisewe did some tests with Marie to make some
vibrations on the door.
below, I exhibit some pictures of
 Accelerometer_on_door_wo_noise.pngAccelerometer_on_door_with_few_noise.pngAccelerometer_on_door_with_few_noise2.png 
  461   Mon Dec 1 18:03:29 2025 Ronic ChicheFixedreportlasers and optics | detectors and electronicsThomX iglooNoise issuesince several days, we see that the locking
is more and more unstable and the lock duration
was sometime less than 20-30 seconds !
 2025-12-01_-_long_run_few_delocks.png 
  462   Wed Dec 3 16:23:40 2025 Ronic ChicheFixedreportlasers and optics | detectors and electronicsThomX iglooNoise issueyesterday, in the same conditions, we had
so many locking issues... exactly
in the same way than before. :-(
 Screenshot_2025-12-03_15-36-59.png 
  463   Mon Dec 8 11:17:17 2025 Ronic ChicheFixedissuelasers and optics | detectors and electronicsThomX iglooHigh frequency noiseThis morning, I installed a high-voltage
probe (1:1000) on the AC-line to see if one
could detect a correlation between lock losses
 AC-line.png 
  464   Mon Feb 2 15:29:45 2026 Ronic ChicheFixedinfolasers and opticsThomX iglooLaser Amplifier used at 70%today, I installed a laser dump just before
the telescope to avoid any high power (50W
instead of 17W usually) issue in the injection
 img1.jpg 
  465   Wed Apr 22 09:52:40 2026 Ronic ChicheUnder Processreportlasers and opticsThomX iglooIncreasing the FP-cavity power Last Friday, we tried to increase the FP-cavity
power by just increasing the amplifier ratio
from 35% to 70%.
 8x 
  466   Mon Apr 27 16:55:15 2026 Ronic ChicheUnder Processreportlasers and opticsThomX iglooIncreasing the FP-cavity power today with Alice and Daniele,

we checked the distance of the
optical setup :
  
  467   Wed Apr 29 09:38:20 2026 Ronic ChicheUnder Processreportlasers and opticsThomX iglooIncreasing the FP-cavity power new power increase after alignment of the
waveplates and iris.

amplifier ratio (%) => FPC cavity
 Measured_data_v2.png 
  468   Wed Apr 29 15:31:06 2026 Ronic ChicheUnder Processreportlasers and opticsThomX iglooIncreasing the FP-cavity power I installed a stronger optical filter on
the reflection photodiode and loaded it on
1kOhms.
 Sans_titre.png 
  469   Mon May 18 11:33:46 2026 Ronic ChicheUnder Processreportlasers and opticsThomX iglooIncreasing the FP-cavity power this morning with Daniele, we set the amplifier
ratio to 33% => 97kW

then, we set the amplifier ratio
  
  470   Thu May 28 10:31:23 2026 Ronic ChicheUnder Processreportlasers and opticsThomX iglooIncreasing the FP-cavity power Last friday, with Daniele, we did some
realignement on the 2nd CVBG of the compressor
: the beam shape seems better, now.
  
  471   Wed Jun 10 16:16:33 2026 Ronic ChicheUnder Processinfomechanics | lasers and optics | vacuumThomX iglooremoving of the M1 mirrorafter trying to increase the power and
after moving S3 to try to change the size
of the beam on the mirrors, the FPC power
  
  472   Thu Jun 11 11:16:55 2026 Ronic ChicheUnder Processinfomechanics | lasers and optics | vacuumThomX iglooremoving of the M1 mirrorthis morning, with Daniele, we took some
images of the M1 mirror HR surface :

1- large image span of the mirror
 6x 
  473   Thu Jun 11 17:11:02 2026 Ronic ChicheUnder Processinfomechanics | lasers and optics | vacuumThomX iglooremoving of the M1 mirrorThis afternoon, we installed a new M1 mirror,
the only one for which the box was not already
opened.
  
  15   Fri Feb 1 14:10:25 2019 Loïc AmoudryFixedreportlasers and opticsOtherMirrors sent to LMA30/01/19 - As LMA people came, we discussed
about the spots on ThomX mirrors which should
be clean. As mentionned in "SBOX commissioning",
  
  40   Tue Jun 11 18:03:10 2019 Loïc AmoudryFixedreportmechanics | lasers and opticsOtherAmplifier mission BordeauxVoyage Bordeaux au CELIA avec Jérôme
LHERMITE (Loïc et Titoutan)

1er jour (34°) :
 1st_and_2nd_stage.jpg3rd_stage_pumps.jpg3rd_stage_signal.jpg 
  41   Wed Jun 12 12:07:11 2019 Loïc AmoudryFixedreportmechanics | lasers and opticsOtherAmplifier mission BordeauxM² mesuré à 10W.
M²x = 1.17
M²y = 1.14
 Ampli_ThomX.PNGM2_ThomX.png 
  1   Tue Sep 25 19:17:43 2018 Loïc AmoudryFixedinfolasers and opticsOptical roomOnefive Origami 33MHz specsPower during ~20h, power after turn on,
spectrum image and excel datas, polarisation
and pulse length.
 6x 
  2   Thu Sep 27 15:43:48 2018 Ronic ChicheFixedinfolasers and opticsOptical roomOnefive Origami 33MHz specs - Polarisationone rotate a lambda/2 before a PBS and
meausure the refected power:

Pmax=35.6 mW
 polar_onefive_33MHz_polarisation.jpg 