257
Thu Aug 27 11:08:44 2026
Ronic Chiche
Fixed
info
lasers and optics
Optical room
fiber patch cables attenuation measurements
attenuation with 1 mating sleeve :
green jacket fiber : -0.4 dB
blue jacket fiber N°1 : -0.5
258
Thu Aug 27 11:13:07 2026
Ronic Chiche
Under Process
info
lasers and optics
Optical room
Menhir oscillator phase noise measurement
with Alice, we installed the Menhir 160MHz
oscillator on a breadboad placed on top of
4 damped feet around the amplifier.
259
Mon Aug 31 15:14:04 2026
Ronic Chiche
Under Process
info
lasers and optics
Optical room
Menhir oscillator phase noise measurement
measurements made with Alice:
comparison between the CW OEwaves
oscillator phase noise measured last week
260
Mon Aug 31 16:15:46 2026
Ronic Chiche
Under Process
info
lasers and optics
Optical room
Menhir oscillator phase noise measurement
comparison of the Menhir 160MHz phase noise
with the CW reference.
- in blue, the oscillators are
261
Mon Aug 31 16:28:31 2026
Ronic Chiche
Under Process
info
lasers and optics
Optical room
Menhir oscillator phase noise measurement
after the previous
measurement, we installed similar dampers,
than the ones used belows the aluminium breadboard,
61
Tue Jan 5 12:38:04 2021
Manar Amer
Fixed
report
lasers and optics
Optical room
Transmission Measurements for the Mirrors.
Thursday 10 December 2020:
First Test Measurement - Inside Box
The
transmission ( T ) for the mirrors of SBOX
6x
64
Mon Jul 5 14:14:01 2021
Manar Amer
Fixed
report
lasers and optics | detectors and electronics
Optical room
OEwaves CW Laser - Phase Noise
We started measuring the phase
noise on the OEwaves CW laser.
Class
3b
wavelength 1.5 um
The
65
Mon Jul 5 14:40:57 2021
Manar Amer
Fixed
report
lasers and optics | detectors and electronics
Optical room
OEwaves CW Laser - Phase Noise
a correction on the wavelength of the laser
it is 1030 um
66
Mon Jul 5 14:46:44 2021
Manar Amer
Fixed
report
lasers and optics | detectors and electronics
Optical room
OEwaves CW Laser - Phase Noise
correction on unit 1030 nm
67
Wed Mar 23 18:11:22 2022
Manar Amer
Fixed
report
utilities
Optical room
Dust Cleaning of SBox
An Initial cleaning of the KBoX outside
was done, All the optics and mechanics placed
that were on the table were removed.
68
Wed Mar 23 18:17:45 2022
Manar Amer
Fixed
report
utilities
Optical room
Dust Cleaning of SBox
The cleaning was done twice with half an
hour interval between the two times
reading on the dust meter was taken
69
Tue Mar 29 16:23:00 2022
Manar Amer
Fixed
report
lasers and optics | cabling
Optical room
Mightylaser Amplifier
Migthylaser amplifier has been moved from
the SBox table to the PLIC table.
70
Tue Mar 29 16:27:34 2022
Manar Amer
Fixed
report
utilities
Optical room
Spincoater - Mirrror cleaning
In preparation for cleaning the mirrors.
The spincoater has been moved with
its bump into the sas of the SBox.
71
Wed Mar 30 18:53:01 2022
Manar Amer
Fixed
report
mechanics | lasers and optics | utilities
Optical room
Dust Cleaning of SBox
The inside of the cavity was cleaned today.
It was very dirty and had big dust particles
and fibers inside.
6x
72
Thu Mar 31 13:29:01 2022
Manar Amer
Fixed
report
mechanics | lasers and optics | utilities
Optical room
Dust Cleaning of SBox
After an overnight where the cavity box
was left open, a measurement of the Dust
meter was taken:
73
Mon Apr 4 13:58:23 2022
Manar Amer
Fixed
issue
detectors and electronics | utilities
Optical room
Clean Airflow Dust measurements
The dust meter cap was cleaned using Alcohol,
and using the filter white cap the dust count
was (0 Av. 10 min)
74
Fri Apr 8 19:27:12 2022
Manar Amer
Fixed
issue
detectors and electronics | utilities
Optical room
Clean Airflow Dust measurements
Dust measurement done today on top
of the SBox average 10min
75
Wed Apr 13 18:42:50 2022
Manar Amer
Fixed
report
lasers and optics | utilities
Optical room
Spincoater - Mirrror cleaning
Used the spincoater and pure water to clean
dielectric mirrors which will be installed
inside the SBox for directing the reflection
76
Wed Apr 20 19:05:37 2022
Manar Amer
Fixed
report
lasers and optics
Optical room
KBox Alignment
Starting with the alignment of the KBox
using CW Koheras.
The size of the CW output from
7x
77
Mon Apr 25 17:28:29 2022
Manar Amer
Fixed
report
lasers and optics
Optical room
SBox Alignment
SBox alignment, using CW laser Koheras
power 1 mW
6x