  Status of commissioning, report also here plots are reports., Page 6 of 6  Not logged in ELOG logo
Entry   Polarization optimization, posted by Loïc Amoudry at Optical room about lasers and optics 
Optimization of the polarization has been made the 03/10/18. Checked in reflection of the cavity in reflection&transmission of a PBS, locked and
unlocked. Only with 2nd stage.

Ratio values are reflection of PBS divided by transmission or the opposite.
Entry   Phase measurement with HASO, posted by Loïc Amoudry at Optical room about lasers and optics 180928_M2_0A.has1801001_input_0A.has180928_M2_trans_0A.txt
2 measurement.

First one locked, in transmission of M2 with 2nd stage 0A.Total 89 cm from the big waist (planar mirrors). 2 wedges used.

Second one at input beam with the same power. Datas taken at the same equivalent position than the first one. 3 wedges used.
    Reply   Phase measurement with HASO, posted by Loïc Amoudry at Optical room about lasers and optics 8x
* Loic has to fix the number of files (3) regarding the number of measurements (2)

* splitted Intensity and phase HASO files

* image for each file
       Reply   Phase measurement with HASO, posted by Loïc Amoudry at Optical room about lasers and optics 
with just intensity, the coupling is 98%, and with phase, x direction coupling is 95%, y direction 97%, so the telescope is good.




Loïc
Amoudry wrote:



* Loic has to fix the number of files (3) regarding the number of
Entry   Cavity polarization states (Koheras), Finesse of 2 polarization states, posted by Loïc Amoudry at Optical room about lasers and optics Finesse_higher.isfFinesse_lower.isfThe2polarizations_states_direct_trans.isfThe2polarizations_states_direct_trans.png
Measurments of Finesse with the 2 polarization states, let's call them H (higher) and L (lower): 24500 for the H and 23500 for the L.

We checked the polarization states in transmission of the FP cavity after a PBS. The H was stronger in PBS trans and the L stronger in PBS ref.

We measured the power in reflection of the PBS and added a half WP that we aligned with the PBS polarization. Then, to get the maximum power