HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports., Page 4 of 13  Not logged in ELOG logo
Entry   Transmission Measurements for the Mirrors. , posted by Manar Amer at Optical room about lasers and optics 6x
  • Thursday 10 December 2020:   First Test Measurement - Inside Box
    •  The transmission ( T ) for the mirrors of SBOX and ThomX was measured over a range from the center of the mirror up to 5mm.
    •  
    • The first setup was initially done while the mirrorr is installed inside the metal box, when measuring T we observe that there is an erregular  change in the power measured as we move azay from the center of the mirror, this was due to reflection of light inside the box. The reflection from the mirror was reflected on the metal inside the box and it was measured by the power meter.

               (so for this setup we only took measrements for T along the center of the mirror)

The file stating the measurements is : Test Cavity _ Mirrors transmission _ Inside box

=======================================================================

  • Tuesday 15 December 2020 :  Second Test - Outside Box
    • The setup for T test was moved outside the metal box with the same distances between the injection beam and mirror ~ 50 cm. and the distance between the injection laser and the mirror next to it ~ 9 cm.
    •  
    • Observation for T is done at full angle and at half angle of reflection ( the mirror was tilted to achive this )
    •  
    • we observed that after each change in the distance from the center the power meter needed to be aligned to have the max power.
    • after doing all the measurements for the range from the center we made it again for the center only and the consistance of the transmission was good.

The file stating the measurements is : Test Cavity _ Mirrors transmission _ outside box

note: the laser used for the test is the Koheras

details on the pwer used, mirrors, power meter, filters, ....etc are shown in the Excel data sheets

Done By : Loic , Manar and Ahmed

Entry   Origami SN2440 133MHz laser power measurment... and failure, posted by Ronic Chiche at Optical room about lasers and optics 133mhzpower.PNG133mhz_power_zoom.png

we received from NKT the Origami SN2440 133MHz oscillator from repair on 2020/12/17
and we measured the power trend from 2020/12/21 until the whole winter Hollyday, during 14 days and 17h (1 273 334 s).
the power was quite constant, about 58mW (the measurement has been done with the OD2 filter and using a (x20) factor in the software to compensate for it),
except for some peaks at the beginning of the measurement to 60.7mW.
I tried to reproduce these peaks by switching on several lights in the room and the airflow ceil but the effect is neglectable,
thus these peaks seem to really come from the oscillator power.

at the beginning of January, the oscillator suffered a back reflection from the rugged anodized convex surface of the power meter (apparently not from the OD2 filter itself) and the laser stopped immediately to modelock.
the laser power dropped to 0 and the laser controller led started to blink red. NKT has been contacted and they think it could be the laser pump diodes that have been damaged.

for the future, we will have to install a fixed optical isolator from the start, even for short operations.

 

Entry   Origami SN2440 133MHz laser tests, posted by Ronic Chiche at Optical room about lasers and optics IMG_20210610_161502.jpgspectre_133MHz.PNGScreen_0015.png

Today, we received the Origami SN2440 133MHz oscillator from repair (NKT mentioned a defective wire inside the controller....)

We immediately installed it behind an isolator (Faraday rotator+PBS+halfwave plate).

the output power is around 60mW.
the spectrum is around 1030nm
the repetition rate is around 133.33MHz

Entry   OEwaves CW Laser - Phase Noise , posted by Manar Amer at Optical room about lasers and optics | detectors and electronics 20210207_OEwaves_laser.jpg20210207_Self_coupling_setup.jpgSchematic-setup-for-optical-delayed-self-coupling-detection.png
  • We started measuring the phase noise on the OEwaves CW laser.
    • Class 3b
    • wavelength 1.5 um
  • The procedure is done using self coupling of the laser
    • splitter 50%-50%\
    • delay line 100 m 
    • all fibers are PM type (polarity maintained)
    • Photodetector is "lab buddy", very fast diode.
    • Note: différance from schematic (we did not use a low pass filter)
    Reply   OEwaves CW Laser - Phase Noise , posted by Manar Amer at Optical room about lasers and optics | detectors and electronics 

a correction on the wavelength of the laser  it is 1030 um 

Manar Amer wrote:
  • We started measuring the phase noise on the OEwaves CW laser.
    • Class 3b
    • wavelength 1.5 um
  • The procedure is done using self coupling of the laser
    • splitter 50%-50%\
    • delay line 100 m 
    • all fibers are PM type (polarity maintained)
    • Photodetector is "lab buddy", very fast diode.
    • Note: différance from schematic (we did not use a low pass filter)

 

    Reply   OEwaves CW Laser - Phase Noise , posted by Manar Amer at Optical room about lasers and optics | detectors and electronics 

correction on unit 1030 nm

Manar Amer wrote:

a correction on the wavelength of the laser  it is 1030 um 

Manar Amer wrote:
  • We started measuring the phase noise on the OEwaves CW laser.
    • Class 3b
    • wavelength 1.5 um
  • The procedure is done using self coupling of the laser
    • splitter 50%-50%\
    • delay line 100 m 
    • all fibers are PM type (polarity maintained)
    • Photodetector is "lab buddy", very fast diode.
    • Note: différance from schematic (we did not use a low pass filter)

 

 

Entry   Dust Cleaning of SBox, posted by Manar Amer at Optical room about utilities Log_Top_window.jpgLog_Reflection_window.jpgLog_Injection_Side_window.jpgLog_Injection_side_window2.jpg

An Initial cleaning of the KBoX outside was done, All the optics and mechanics placed that were on the table were removed.

Only the old Mitylaser amplifier remains on the table.

Since we will be using the ThomX Celia amplifier, we will remove it and place it in a secure location (to be decided).

Tomorrow a cleaning of the inside of the cavity box is to be done.

    Reply   Dust Cleaning of SBox, posted by Manar Amer at Optical room about utilities 

The cleaning was done twice with half an hour interval between the two times
reading on the dust meter was taken after each time, and it was zero averaged over a 10-min

Manar Amer wrote:

An Initial cleaning of the KBoX outside was done, All the optics and mechanics placed that were on the table were removed.

Only the old Mitylaser amplifier remains on the table.

Since we will be using the ThomX Celia amplifier, we will remove it and place it in a secure location (to be decided).

Tomorrow a cleaning of the inside of the cavity box is to be done.

 

Entry   Mightylaser Amplifier , posted by Manar Amer at Optical room about lasers and optics | cabling 

Migthylaser amplifier has been moved from the SBox table to the PLIC table.

 

Entry   Spincoater - Mirrror cleaning , posted by Manar Amer at Optical room about utilities 

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

    Reply   Dust Cleaning of SBox, posted by Manar Amer at Optical room about mechanics | lasers and optics | utilities 6x

The inside of the cavity was cleaned today. It was very dirty and had big dust particles and fibers inside.

Cleaning was done for everything; wall, floor, fixed mounts, wires (still have some dust difficult to remove), edges.

The dust meter average was measured outside the cavity (2471 p/m3 av 6 min)  and inside after cleaning (1424 p/m3 av 10 min).

The cavity was left open under the clean air flow, will observe it under UV, clean and measure Dust tomorrow.

Images attached are after cleaning.

Objects inside the cavity at the moment :

  • 4 main mirror mounts
  • 4 mirror fixers on mounts and 3 screws each (total 12)
  • 2 reflecting mirror mounts
  • 2 dielectric mirrors (mounted and to be cleaned)
  • two mounts for D-shaped mirrors.
  • 2 D-shaped mirrors (unmounted, and they will need cleaning)
  • 2 metal irises (pinholes)

 

Manar Amer wrote:

The cleaning was done twice with half an hour interval between the two times
reading on the dust meter was taken after each time, and it was zero averaged over a 10-min

Manar Amer wrote:

An Initial cleaning of the KBoX outside was done, All the optics and mechanics placed that were on the table were removed.

Only the old Mitylaser amplifier remains on the table.

Since we will be using the ThomX Celia amplifier, we will remove it and place it in a secure location (to be decided).

Tomorrow a cleaning of the inside of the cavity box is to be done.

 

 

    Reply   Dust Cleaning of SBox, posted by Manar Amer at Optical room about mechanics | lasers and optics | utilities 20220331_Dust_Meter_Outside_cavity_Av10min.jpg20220331_Dust_Meter_Inside_cavity_Av10min.jpg

After an overnight where the cavity box was left open, a measurement of the Dust meter was taken:

Outside on the optical table on direct air flux (1701 p/m3 Av 10 min) and

Inside the cavity box (998 p/m3 Av 10 min)

The dust inside the cavity has decreased by ~ 30 % for 0.3 u particles.

Note: the cavity box is still open.

Manar Amer wrote:

The inside of the cavity was cleaned today. It was very dirty and had big dust particles and fibers inside.

Cleaning was done for everything; wall, floor, fixed mounts, wires (still have some dust difficult to remove), edges.

The dust meter average was measured outside the cavity (2471 p/m3 av 6 min)  and inside after cleaning (1424 p/m3 av 10 min).

The cavity was left open under the clean air flow, will observe it under UV, clean and measure Dust tomorrow.

Images attached are after cleaning.

Objects inside the cavity at the moment :

  • 4 main mirror mounts
  • 4 mirror fixers on mounts and 3 screws each (total 12)
  • 2 reflecting mirror mounts
  • 2 dielectric mirrors (mounted and to be cleaned)
  • two mounts for D-shaped mirrors.
  • 2 D-shaped mirrors (unmounted, and they will need cleaning)
  • 2 metal irises (pinholes)

 

Manar Amer wrote:

The cleaning was done twice with half an hour interval between the two times
reading on the dust meter was taken after each time, and it was zero averaged over a 10-min

Manar Amer wrote:

An Initial cleaning of the KBoX outside was done, All the optics and mechanics placed that were on the table were removed.

Only the old Mitylaser amplifier remains on the table.

Since we will be using the ThomX Celia amplifier, we will remove it and place it in a secure location (to be decided).

Tomorrow a cleaning of the inside of the cavity box is to be done.

 

 

 

Entry   Clean Airflow Dust measurements, posted by Manar Amer at Optical room about detectors and electronics | utilities 

The dust meter cap was cleaned using Alcohol, and using the filter white cap the dust count was (0 Av. 10 min)

After the dust counter was hand held in direct airflow, it counted (1581 p/m3 Av. 2 min).

I tested the count also in the SAS, and it counted twice the amount ~ 4500 p/m3

Note: discussion on the next steps to take for the airflow filtering !!!!!!!!

    Reply   Clean Airflow Dust measurements, posted by Manar Amer at Optical room about detectors and electronics | utilities 20220408_DustMeasurement.jpg

Dust measurement done today on  top of the SBox average 10min

Manar Amer wrote:

The dust meter cap was cleaned using Alcohol, and using the filter white cap the dust count was (0 Av. 10 min)

After the dust counter was hand held in direct airflow, it counted (1581 p/m3 Av. 2 min).

I tested the count also in the SAS, and it counted twice the amount ~ 4500 p/m3

Note: discussion on the next steps to take for the airflow filtering !!!!!!!!

 

    Reply   Spincoater - Mirrror cleaning , posted by Manar Amer at Optical room about lasers and optics | utilities Reflection_one_front_after_cleaning.jpgreflection_two_front_before_cleaning.jpgReflection_two_front_after_cleaning.jpg

Used the spincoater and pure water to clean dielectric mirrors which will be installed inside the SBox for directing the reflection line.

The method : place pure alchole on mirror - wipe it using white sterile tissue (in one direction) - clean with pure water in spincoater suing cotton swap .

After multible rounds, the result in the images was obtained. there is still a significant amount of dust particles (not all could be dust!!).

Not sure if the reason is the cleaning method ?? will invistegate it more !!!

Manar Amer wrote:

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

 

Entry   KBox Alignment , posted by Manar Amer at Optical room about lasers and optics 7x

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

    Reply   SBox Alignment , posted by Manar Amer at Optical room about lasers and optics 6x

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

    Reply   SBox Alignment , posted by Manar Amer at Optical room about lasers and optics Ma1_Ma2_Horizontal_Position_Optimized.pngMa1_Ma2_Vertical_Position_Optimized.pngAfter_M1_diaphram_1m_Saturated_Beam_Optimized.jpgAfter_M2_M1_diaphram_Saturated_Beam_Optimized.jpgAfter_M2_M1_diaphram_Beam_Optimized.jpg

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

    Reply   SBox Alignment , posted by Manar Amer at Optical room about lasers and optics 

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

    Reply   SBox Alignment , posted by Manar Amer at Optical room about lasers and optics 9x
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

ELOG V3.1.4-395e101