HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports., Page 4 of 13  Not logged in ELOG logo
ID Date Author Status Type Category Location Titledown
  155   Fri Sep 9 15:47:46 2022 Manar AmerFixedreportlasers and opticsOptical roomPower storage In Cavity

At transmission power of 42 mW, coupling 60%, we see fundamental mode with a degeneracy

The D-shape motors were moved to a position 2 000 000 steps (in theory they should be the max position)

but no change appeared on the 00 mode or the higher order mode.

We will break the vacuum and check the position of the D-shaped mirrors.

Manar Amer wrote:

We start of power output from amplifier of 1W with only the first and second stage on then we start with the third stage power increase,

Note: when changing the current on the third stage 4 diodes, better to do it step by step for each one with a step of ~ 0.5 A

  •  

Amplifier output current (A)

for four diodes

Power Output (W)

Transmitted power (MW)

 

0 , 0 , 0 ,  0 ~ 1 2.3
1 , 1 , 1 , 1 / 11.0
1.5 , 1.3, 1 , 1 / 35
1.5 , 1.5 , 1 , 1 / 42.16
   

 

  • we placed at the transmission point a splitter the transmitted power before was 35 mW after placing it we had 26 mW
    • 97 % transmission , 3% reflected which goes to the beam profiler

 

  • We see degeneracy modes at the last step of 42.16 mW , so we started to test the D-shaped mirrors
    • the full range of the motors is 50.8 mm
    • each step =  30 nm , 30 000 steps = 1 mm

 

  158   Mon Sep 12 16:27:14 2022 Manar AmerFixedreportlasers and opticsOptical roomPower storage In Cavity

Power increased in the amplifier

reflective filters added at the transmission point to be able to have signal on the Beam Profiler, Transmission diode and power meter

Amplifier output current (A)

diode 1

Amplifier output current (A)

diode 2

Amplifier output current (A)

diode 3

Amplifier output current (A)

diode 4

Power Output (W) Transmitted power (mW) note
0.5 0.5 0.5 0.5     no change in the beam shape
1 1 1 1      
1.5 1 1 1     Appearance of higher order mode (vertical), suppressed using D-shaped mirror (vertical only) by (-23000), image recorded.
1.5 1.5 1.5 1.5     no change
2 1.8

1.5

1.5   52 mA 52 mA after CEP adjustment , power measured after 2 reflective fillers NDUV30A and NDUV20
2 2 2 2   70 mA  
2.5 2.5 2 2   95  
2.5 2.5 2.5 2.5    

sudden drop in power, cavity lock lost, lock system looks ok (maybe something in the cavity)

mode shape is the same

coupling is large, transmission is very low

change the CEP, mo effect in changing it (we have no gain in the cavity, No Finesse)  could be something happened to the mirrors !!!!!!

 

 

Manar Amer wrote:

At transmission power of 42 mW, coupling 60%, we see fundamental mode with a degeneracy

The D-shape motors were moved to a position 2 000 000 steps (in theory they should be the max position)

but no change appeared on the 00 mode or the higher order mode.

We will break the vacuum and check the position of the D-shaped mirrors.

Manar Amer wrote:

We start of power output from amplifier of 1W with only the first and second stage on then we start with the third stage power increase,

Note: when changing the current on the third stage 4 diodes, better to do it step by step for each one with a step of ~ 0.5 A

  •  

Amplifier output current (A)

for four diodes

Power Output (W)

Transmitted power (MW)

 

0 , 0 , 0 ,  0 ~ 1 2.3
1 , 1 , 1 , 1 / 11.0
1.5 , 1.3, 1 , 1 / 35
1.5 , 1.5 , 1 , 1 / 42.16
   

 

  • we placed at the transmission point a splitter the transmitted power before was 35 mW after placing it we had 26 mW
    • 97 % transmission , 3% reflected which goes to the beam profiler

 

  • We see degeneracy modes at the last step of 42.16 mW , so we started to test the D-shaped mirrors
    • the full range of the motors is 50.8 mm
    • each step =  30 nm , 30 000 steps = 1 mm

 

 

  159   Mon Sep 12 17:00:35 2022 Manar AmerFixedreportlasers and opticsOptical roomPower storage In Cavity

at max stage we reached at 2.5 A on the diodes we have

8000 gain , coupling ~ 60% , with power inside the cavity 50 kW

 

The higer order mode that was observed today,

it was also observed on Friday  along the other higher order mode recorded on the logbook.

after cutting it with the D-shaped mirror we recorded the shape (image attached)

Manar Amer wrote:

Power increased in the amplifier

reflective filters added at the transmission point to be able to have signal on the Beam Profiler, Transmission diode and power meter

Amplifier output current (A)

diode 1

Amplifier output current (A)

diode 2

Amplifier output current (A)

diode 3

Amplifier output current (A)

diode 4

Power Output (W) Transmitted power (mW) note
0.5 0.5 0.5 0.5     no change in the beam shape
1 1 1 1      
1.5 1 1 1     Appearance of higher order mode (vertical), suppressed using D-shaped mirror (vertical only) by (-23000), image recorded.
1.5 1.5 1.5 1.5     no change
2 1.8

1.5

1.5   52 mA 52 mA after CEP adjustment , power measured after 2 reflective fillers NDUV30A and NDUV20
2 2 2 2   70 mA  
2.5 2.5 2 2   95  
2.5 2.5 2.5 2.5    

sudden drop in power, cavity lock lost, lock system looks ok (maybe something in the cavity)

mode shape is the same

coupling is large, transmission is very low

change the CEP, mo effect in changing it (we have no gain in the cavity, No Finesse)  could be something happened to the mirrors !!!!!!

 

 

Manar Amer wrote:

At transmission power of 42 mW, coupling 60%, we see fundamental mode with a degeneracy

The D-shape motors were moved to a position 2 000 000 steps (in theory they should be the max position)

but no change appeared on the 00 mode or the higher order mode.

We will break the vacuum and check the position of the D-shaped mirrors.

Manar Amer wrote:

We start of power output from amplifier of 1W with only the first and second stage on then we start with the third stage power increase,

Note: when changing the current on the third stage 4 diodes, better to do it step by step for each one with a step of ~ 0.5 A

  •  

Amplifier output current (A)

for four diodes

Power Output (W)

Transmitted power (MW)

 

0 , 0 , 0 ,  0 ~ 1 2.3
1 , 1 , 1 , 1 / 11.0
1.5 , 1.3, 1 , 1 / 35
1.5 , 1.5 , 1 , 1 / 42.16
   

 

  • we placed at the transmission point a splitter the transmitted power before was 35 mW after placing it we had 26 mW
    • 97 % transmission , 3% reflected which goes to the beam profiler

 

  • We see degeneracy modes at the last step of 42.16 mW , so we started to test the D-shaped mirrors
    • the full range of the motors is 50.8 mm
    • each step =  30 nm , 30 000 steps = 1 mm

 

 

 

  164   Wed Sep 14 16:16:08 2022 Manar AmerFixedreportlasers and opticsOptical roomPower storage In Cavity

The cavity was locked under vacuum, coupling is approximately < 10%

The lock was relatively easy and the usual sign (oscillations on the transmission tail) of high finesse on the transmission signal was not seen.

 

Manar Amer wrote:

After changing the mirror M1 and cleaning the 2 mirrors

They were placed in the cavity box (avoiding touching the mounts to not affect the alignment and placing the mirrors as close to the previous position)

we saw horizontal higher order modes immediately after injecting power,  horizontal misalignment!!.

We aligned the injected beam, adjusted the cavity length and saw the modes 01 (high transition) and 00 (ok but still much lower than 01)

The image attached shows some coupling < 5% , we adjusted the CEP, but it was the max coupling.

we will need to align better tomorrow to increase 00 modes.

The cavity box is closed and placed under vacuum again.

 

 

Manar Amer wrote:

at max stage we reached at 2.5 A on the diodes we have

8000 gain , coupling ~ 60% , with power inside the cavity 50 kW

 

The higer order mode that was observed today,

it was also observed on Friday  along the other higher order mode recorded on the logbook.

after cutting it with the D-shaped mirror we recorded the shape (image attached)

Manar Amer wrote:

Power increased in the amplifier

reflective filters added at the transmission point to be able to have signal on the Beam Profiler, Transmission diode and power meter

Amplifier output current (A)

diode 1

Amplifier output current (A)

diode 2

Amplifier output current (A)

diode 3

Amplifier output current (A)

diode 4

Power Output (W) Transmitted power (mW) note
0.5 0.5 0.5 0.5     no change in the beam shape
1 1 1 1      
1.5 1 1 1     Appearance of higher order mode (vertical), suppressed using D-shaped mirror (vertical only) by (-23000), image recorded.
1.5 1.5 1.5 1.5     no change
2 1.8

1.5

1.5   52 mA 52 mA after CEP adjustment , power measured after 2 reflective fillers NDUV30A and NDUV20
2 2 2 2   70 mA  
2.5 2.5 2 2   95  
2.5 2.5 2.5 2.5    

sudden drop in power, cavity lock lost, lock system looks ok (maybe something in the cavity)

mode shape is the same

coupling is large, transmission is very low

change the CEP, mo effect in changing it (we have no gain in the cavity, No Finesse)  could be something happened to the mirrors !!!!!!

 

 

Manar Amer wrote:

At transmission power of 42 mW, coupling 60%, we see fundamental mode with a degeneracy

The D-shape motors were moved to a position 2 000 000 steps (in theory they should be the max position)

but no change appeared on the 00 mode or the higher order mode.

We will break the vacuum and check the position of the D-shaped mirrors.

Manar Amer wrote:

We start of power output from amplifier of 1W with only the first and second stage on then we start with the third stage power increase,

Note: when changing the current on the third stage 4 diodes, better to do it step by step for each one with a step of ~ 0.5 A

  •  

Amplifier output current (A)

for four diodes

Power Output (W)

Transmitted power (MW)

 

0 , 0 , 0 ,  0 ~ 1 2.3
1 , 1 , 1 , 1 / 11.0
1.5 , 1.3, 1 , 1 / 35
1.5 , 1.5 , 1 , 1 / 42.16
   

 

  • we placed at the transmission point a splitter the transmitted power before was 35 mW after placing it we had 26 mW
    • 97 % transmission , 3% reflected which goes to the beam profiler

 

  • We see degeneracy modes at the last step of 42.16 mW , so we started to test the D-shaped mirrors
    • the full range of the motors is 50.8 mm
    • each step =  30 nm , 30 000 steps = 1 mm

 

 

 

 

 

  165   Wed Sep 14 17:21:59 2022 Manar AmerFixedreportlasers and opticsOptical roomPower storage In Cavity

We improved the lock using the alignment, polarization and CEP , the maximum we got was ~ 12% coupling

only the second stage is on !

Decision : this is the maximum we can obtain using the M1 from mighty laser

tomorrow we break the vacuum and try with shifting M1 from ThomX position to avoid the damaged spot

Manar Amer wrote:

The cavity was locked under vacuum, coupling is approximately < 10%

The lock was relatively easy and the usual sign (oscillations on the transmission tail) of high finesse on the transmission signal was not seen.

 

Manar Amer wrote:

After changing the mirror M1 and cleaning the 2 mirrors

They were placed in the cavity box (avoiding touching the mounts to not affect the alignment and placing the mirrors as close to the previous position)

we saw horizontal higher order modes immediately after injecting power,  horizontal misalignment!!.

We aligned the injected beam, adjusted the cavity length and saw the modes 01 (high transition) and 00 (ok but still much lower than 01)

The image attached shows some coupling < 5% , we adjusted the CEP, but it was the max coupling.

we will need to align better tomorrow to increase 00 modes.

The cavity box is closed and placed under vacuum again.

 

 

Manar Amer wrote:

at max stage we reached at 2.5 A on the diodes we have

8000 gain , coupling ~ 60% , with power inside the cavity 50 kW

 

The higer order mode that was observed today,

it was also observed on Friday  along the other higher order mode recorded on the logbook.

after cutting it with the D-shaped mirror we recorded the shape (image attached)

Manar Amer wrote:

Power increased in the amplifier

reflective filters added at the transmission point to be able to have signal on the Beam Profiler, Transmission diode and power meter

Amplifier output current (A)

diode 1

Amplifier output current (A)

diode 2

Amplifier output current (A)

diode 3

Amplifier output current (A)

diode 4

Power Output (W) Transmitted power (mW) note
0.5 0.5 0.5 0.5     no change in the beam shape
1 1 1 1      
1.5 1 1 1     Appearance of higher order mode (vertical), suppressed using D-shaped mirror (vertical only) by (-23000), image recorded.
1.5 1.5 1.5 1.5     no change
2 1.8

1.5

1.5   52 mA 52 mA after CEP adjustment , power measured after 2 reflective fillers NDUV30A and NDUV20
2 2 2 2   70 mA  
2.5 2.5 2 2   95  
2.5 2.5 2.5 2.5    

sudden drop in power, cavity lock lost, lock system looks ok (maybe something in the cavity)

mode shape is the same

coupling is large, transmission is very low

change the CEP, mo effect in changing it (we have no gain in the cavity, No Finesse)  could be something happened to the mirrors !!!!!!

 

 

Manar Amer wrote:

At transmission power of 42 mW, coupling 60%, we see fundamental mode with a degeneracy

The D-shape motors were moved to a position 2 000 000 steps (in theory they should be the max position)

but no change appeared on the 00 mode or the higher order mode.

We will break the vacuum and check the position of the D-shaped mirrors.

Manar Amer wrote:

We start of power output from amplifier of 1W with only the first and second stage on then we start with the third stage power increase,

Note: when changing the current on the third stage 4 diodes, better to do it step by step for each one with a step of ~ 0.5 A

  •  

Amplifier output current (A)

for four diodes

Power Output (W)

Transmitted power (MW)

 

0 , 0 , 0 ,  0 ~ 1 2.3
1 , 1 , 1 , 1 / 11.0
1.5 , 1.3, 1 , 1 / 35
1.5 , 1.5 , 1 , 1 / 42.16
   

 

  • we placed at the transmission point a splitter the transmitted power before was 35 mW after placing it we had 26 mW
    • 97 % transmission , 3% reflected which goes to the beam profiler

 

  • We see degeneracy modes at the last step of 42.16 mW , so we started to test the D-shaped mirrors
    • the full range of the motors is 50.8 mm
    • each step =  30 nm , 30 000 steps = 1 mm

 

 

 

 

 

 

  11   Wed Oct 31 11:27:49 2018 Loïc AmoudryFixedreportmechanics | lasers and opticsOptical roomPower measurement with D-shape

Measurements of lot of points with D-shape mirrors well positionned.

Power not optimized to the best but almost. (@4A could have 350 mW).

I (A) Ptrans (mW)  Coupling (%)

0

8 62
1 18 67
1.3 43 72
1.6 76 72
1.9 112 72
2.2 145 72
2.5 177 72
2.8 217 72
3.1 253 72
3.4 281 72
3.7 300 72
4 323 71
4.3 249 71
4.6 379 68
4.9 402 68
5.2 417 67
5.5 435 67
5.8 441 65

 

  5   Tue Oct 9 10:15:17 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomPolarization optimization

Optimization of the polarization has been made the 03/10/18. Checked in reflection of the cavity in reflection&transmission of a PBS, locked and unlocked. Only with 2nd stage.

Ratio values are reflection of PBS divided by transmission or the opposite.

  Unlock value Ratio min/max Lock value Ratio min/max
No optimization        
Reflection 1.21 X 3.89 X
Transmission 14.3 8.5 % 3.1 80 %
Only Lambda/2        
Reflection 4.4 X 2.91 X
Transmission 11 40 % 2.79 96 %
2xLambda/2 + 1Lambda/4        
Reflection 8.3 X 3.4 X
Transmission 6.8 82 % 2.2 65 %
Same + PID optimization        
Reflection 9 X 3.35 X
Transmission 5.76 64 %  2 60 %

 

  202   Thu Mar 21 17:32:59 2024 Xinyi LuFixedinfolasers and opticsOptical roomPolarization issue & 50kW with 22W injection

Today, Ronic and I lock the cavity on the other polarization and achieved 50kW with 22W injection.

- After we optimize and lock the cavity in vertical polarization, we rotate the waveplate at the transmission to minimize the signal, then rotate the waveplate at the injection laser to maximize the signal. We lock the cavity in this horizontal polarization and optimize CEP and alignment. The results are: when injected at 10W, the circulating power in vertical polarization is 21.4kW and in horizontal polarization is 23.3kW. The coupling are both ~30%.

- The cavity reflection signal obtained in horizontal polarization is weaker than that in vertical polarization (1/10), so we usually lock on vertical polarization first.

- We then increase the power to see the change in coupling. From current 2A to 3A, coupling change from 30% to 40%. Finally, we obtained a circulating power of 50kW with 22W injection (3A current). In the initial stage of locking, high-order modes appear, but in stable locking, there is only fundamental mode and no mode degeneration. Although there are many fluctuations in transmission and reflection.

- Tomorrow, we will optimize the coupling and add removable stages under the telescope.

Xinyi Lu wrote:

just to add some details :

about the S and P polarization frequency shift:
the PZT scan is 10Vpp at 3.1Hz => the slope is 62 V/s because of the triangle shape of the PZT scan.
so 100µs of separation of the 2 polarization is equivalent to 6.2mV on the PZT.
as the PZT sensitivity is 3.7Hz/V on Frep, the separation of the 2 polarization is equivalent to 23mHz on Frep.
△Frep/Frep = △v/v =>  △v = 41.8 kHz

about the possibility to separate S and P polarization states on secondary resonances:
the total spectral width is ~2nm which is equivalent to 565GHz and contain about 3500 laser harmonics at 160MHz.
the central spectral harmonic is roughly the number n0=1.82M, so with the first secondary resonance condition, the Frep/FSR detuning corresponds to n0*Frep = (n0+1) FSR
so (Frep - FSR) = FSR/n0 ~ 88Hz.
then if the central frequency, related to n0, is on an S-polarization resonance, the harmonics at (n0+475) will be on the P-polarization resonance and so on...
the conclusion is the power detected by a photodiode on secondary resonances are a mix of S and P polarizations (if the laser input beam is also a combination of S and P polarizations)
and we cannot make an observation of different peaks with different polarizations in transmission of the FP-cavity.
for that, it is mandatory to be on the main resonance with Frep = FSR (CEP~0) as condition of resonance.

Xinyi Lu wrote:

These days, Ronic, Aurélien, Fatematuj and I have been doing some tests on polarization issue, trying to see if it is possible to obtain higher gain under other polarization conditions.

  • We installed an additional half-wave plate + PBS + PD at the transmission. By rotating the waveplate of the injection laser, we can compare the resonance signals of single and full polarization. Figures 1 and 2 demonstrate this comparison. The yellow curve is full polarization and the green one is single polarization. The intensity ratio of the different polarizations is unstable in the open-loop state.
  • Based on PZT scan frequency = 3.1 Hz, amplitude = 10 V, sensitivity = 3.7 Hz/V, time difference between two polarization peaks = 100 us, we can calculate △Frep = 10mHz and △v = 42kHz, which means the frequency variation between two polarizations. We see two polarizations only at the main resonance.
  • By the way, we found two spots behind the M3 window (see Figure 3) and the power of both is related to the intra-cavity power. We moved the position of D-shaped mirror and the second spot became weaker and larger like mirror's edge. Maybe the D-shaped mirror is causing a part of laser to be reflected through the window, but it's unclear exactly how the optical path works.

Next steps:

  • We will lock the cavity in different polarization and see if there is higher gain.
  • We will move D-shaped to the maximum and see if the second spot disappears.
  • We will check the coupling value and try to optimize the telescope using adjustable stages.
Xinyi Lu wrote:

Today, Ronic and I optimized the locking at the amplifier current of 2 A and obtained ~ 21 kW inside the cavity.

- When all the iris open, the injected power is 10 W and the coupling is ~ 40%, corresponding to an effective gain of 2,100 and a full gain of 5,250. But the coupling may not be the true value because there is a large spot around the output beam.

- We have optimized the CEP, alignment, D-shaped mirrors and locking state. We optimized alignment after leaving the iris open and the inside power went from 14kW to 21kW.

- The transmission and reflection signals both have some same fluctuations, and they seem to come from the cavity. It's possible that the over-angled mirror mount could be the cause, but not sure. We will check in different power and see the stability of the signal.

- In addition, we found that the design values of the mirror incidence angles for the SBOX (3.359°, 5.900°) are different from the mirror ratings (1.146°). This may result in parameters such as reflection and transmission being different from the datasheet. It will also change the estimated maximum finesse, gain, and power inside the cavity. It might be better if the mirror parameters could be recalculated based on the actual angle of incidence.

Xinyi Lu wrote:

- Today, Ronic and I locked at the amplifier current of 2 A and obtained ~60% coupling after optimizing the CEP (see Figure 1).

- The injected power is 10 W at 2 A. We measured only 14 kW inside the cavity, which corresponds to an effective gain of 1,400 and a full gain of 2,300. The cavity finesse is 23,000 and the normal gain should be around 6,200.

- We found fluctuations in transmission, possibly because of mode degradation. Tomorrow we will use D-shape mirrors to suppress high-order modes and optimize alignment and locking.

Xinyi Lu wrote:

Today, Ronic and I installed the new telescope and locked the cavity.

- We locked at the amplifier current of 1 A and obtained 32% of coupling. (see Figure 1)

- The telescope was designed for a current of 2 A (output power ~10 W). To inject this power, we need to add some filters to devices.

- For CEP tuning, when we changed the AOM frequency while cavity locking, sometimes it caused unlock and power drops. It will be dangerous in high-power cases. So it's better to optimize the AOM frequency in low power and just tune the laser current in high power. Now the current variation range of the menhir laser is 750mA to 950mA.

Xinyi Lu wrote:

- These days, Ronic, Fatematuj and I measured the beam parameters of the output of the third-stage amplifier.

- We used 2 wedges and reflection filters to reduce the intensity on the CCD.

- We measured multiple points at pump current of 2 A (output power ~10 W). The waist diameter of the output is w_x = 792.26 um, w_y=873.90 um.

- The next step is to design the telescope and improve the coupling efficiency.

 

 

 

 

 

 

  200   Wed Mar 20 18:22:53 2024 Xinyi LuFixedinfolasers and opticsOptical roomPolarization issue

These days, Ronic, Aurélien, Fatematuj and I have been doing some tests on polarization issue, trying to see if it is possible to obtain higher gain under other polarization conditions.

  • We installed an additional half-wave plate + PBS + PD at the transmission. By rotating the waveplate of the injection laser, we can compare the resonance signals of single and full polarization. Figures 1 and 2 demonstrate this comparison. The yellow curve is full polarization and the green one is single polarization. The intensity ratio of the different polarizations is unstable in the open-loop state.
  • Based on PZT scan frequency = 3.1 Hz, amplitude = 10 V, sensitivity = 3.7 Hz/V, time difference between two polarization peaks = 100 us, we can calculate △Frep = 10mHz and △v = 42kHz, which means the frequency variation between two polarizations. We see two polarizations only at the main resonance.
  • By the way, we found two spots behind the M3 window (see Figure 3) and the power of both is related to the intra-cavity power. We moved the position of D-shaped mirror and the second spot became weaker and larger like mirror's edge. Maybe the D-shaped mirror is causing a part of laser to be reflected through the window, but it's unclear exactly how the optical path works.

Next steps:

  • We will lock the cavity in different polarization and see if there is higher gain.
  • We will move D-shaped to the maximum and see if the second spot disappears.
  • We will check the coupling value and try to optimize the telescope using adjustable stages.
Xinyi Lu wrote:

Today, Ronic and I optimized the locking at the amplifier current of 2 A and obtained ~ 21 kW inside the cavity.

- When all the iris open, the injected power is 10 W and the coupling is ~ 40%, corresponding to an effective gain of 2,100 and a full gain of 5,250. But the coupling may not be the true value because there is a large spot around the output beam.

- We have optimized the CEP, alignment, D-shaped mirrors and locking state. We optimized alignment after leaving the iris open and the inside power went from 14kW to 21kW.

- The transmission and reflection signals both have some same fluctuations, and they seem to come from the cavity. It's possible that the over-angled mirror mount could be the cause, but not sure. We will check in different power and see the stability of the signal.

- In addition, we found that the design values of the mirror incidence angles for the SBOX (3.359°, 5.900°) are different from the mirror ratings (1.146°). This may result in parameters such as reflection and transmission being different from the datasheet. It will also change the estimated maximum finesse, gain, and power inside the cavity. It might be better if the mirror parameters could be recalculated based on the actual angle of incidence.

Xinyi Lu wrote:

- Today, Ronic and I locked at the amplifier current of 2 A and obtained ~60% coupling after optimizing the CEP (see Figure 1).

- The injected power is 10 W at 2 A. We measured only 14 kW inside the cavity, which corresponds to an effective gain of 1,400 and a full gain of 2,300. The cavity finesse is 23,000 and the normal gain should be around 6,200.

- We found fluctuations in transmission, possibly because of mode degradation. Tomorrow we will use D-shape mirrors to suppress high-order modes and optimize alignment and locking.

Xinyi Lu wrote:

Today, Ronic and I installed the new telescope and locked the cavity.

- We locked at the amplifier current of 1 A and obtained 32% of coupling. (see Figure 1)

- The telescope was designed for a current of 2 A (output power ~10 W). To inject this power, we need to add some filters to devices.

- For CEP tuning, when we changed the AOM frequency while cavity locking, sometimes it caused unlock and power drops. It will be dangerous in high-power cases. So it's better to optimize the AOM frequency in low power and just tune the laser current in high power. Now the current variation range of the menhir laser is 750mA to 950mA.

Xinyi Lu wrote:

- These days, Ronic, Fatematuj and I measured the beam parameters of the output of the third-stage amplifier.

- We used 2 wedges and reflection filters to reduce the intensity on the CCD.

- We measured multiple points at pump current of 2 A (output power ~10 W). The waist diameter of the output is w_x = 792.26 um, w_y=873.90 um.

- The next step is to design the telescope and improve the coupling efficiency.

 

 

 

 

  201   Thu Mar 21 09:26:50 2024 Xinyi LuFixedinfolasers and opticsOptical roomPolarization issue

just to add some details :

about the S and P polarization frequency shift:
the PZT scan is 10Vpp at 3.1Hz => the slope is 62 V/s because of the triangle shape of the PZT scan.
so 100µs of separation of the 2 polarization is equivalent to 6.2mV on the PZT.
as the PZT sensitivity is 3.7Hz/V on Frep, the separation of the 2 polarization is equivalent to 23mHz on Frep.
△Frep/Frep = △v/v =>  △v = 41.8 kHz

about the possibility to separate S and P polarization states on secondary resonances:
the total spectral width is ~2nm which is equivalent to 565GHz and contain about 3500 laser harmonics at 160MHz.
the central spectral harmonic is roughly the number n0=1.82M, so with the first secondary resonance condition, the Frep/FSR detuning corresponds to n0*Frep = (n0+1) FSR
so (Frep - FSR) = FSR/n0 ~ 88Hz.
then if the central frequency, related to n0, is on an S-polarization resonance, the harmonics at (n0+475) will be on the P-polarization resonance and so on...
the conclusion is the power detected by a photodiode on secondary resonances are a mix of S and P polarizations (if the laser input beam is also a combination of S and P polarizations)
and we cannot make an observation of different peaks with different polarizations in transmission of the FP-cavity.
for that, it is mandatory to be on the main resonance with Frep = FSR (CEP~0) as condition of resonance.

Xinyi Lu wrote:

These days, Ronic, Aurélien, Fatematuj and I have been doing some tests on polarization issue, trying to see if it is possible to obtain higher gain under other polarization conditions.

  • We installed an additional half-wave plate + PBS + PD at the transmission. By rotating the waveplate of the injection laser, we can compare the resonance signals of single and full polarization. Figures 1 and 2 demonstrate this comparison. The yellow curve is full polarization and the green one is single polarization. The intensity ratio of the different polarizations is unstable in the open-loop state.
  • Based on PZT scan frequency = 3.1 Hz, amplitude = 10 V, sensitivity = 3.7 Hz/V, time difference between two polarization peaks = 100 us, we can calculate △Frep = 10mHz and △v = 42kHz, which means the frequency variation between two polarizations. We see two polarizations only at the main resonance.
  • By the way, we found two spots behind the M3 window (see Figure 3) and the power of both is related to the intra-cavity power. We moved the position of D-shaped mirror and the second spot became weaker and larger like mirror's edge. Maybe the D-shaped mirror is causing a part of laser to be reflected through the window, but it's unclear exactly how the optical path works.

Next steps:

  • We will lock the cavity in different polarization and see if there is higher gain.
  • We will move D-shaped to the maximum and see if the second spot disappears.
  • We will check the coupling value and try to optimize the telescope using adjustable stages.
Xinyi Lu wrote:

Today, Ronic and I optimized the locking at the amplifier current of 2 A and obtained ~ 21 kW inside the cavity.

- When all the iris open, the injected power is 10 W and the coupling is ~ 40%, corresponding to an effective gain of 2,100 and a full gain of 5,250. But the coupling may not be the true value because there is a large spot around the output beam.

- We have optimized the CEP, alignment, D-shaped mirrors and locking state. We optimized alignment after leaving the iris open and the inside power went from 14kW to 21kW.

- The transmission and reflection signals both have some same fluctuations, and they seem to come from the cavity. It's possible that the over-angled mirror mount could be the cause, but not sure. We will check in different power and see the stability of the signal.

- In addition, we found that the design values of the mirror incidence angles for the SBOX (3.359°, 5.900°) are different from the mirror ratings (1.146°). This may result in parameters such as reflection and transmission being different from the datasheet. It will also change the estimated maximum finesse, gain, and power inside the cavity. It might be better if the mirror parameters could be recalculated based on the actual angle of incidence.

Xinyi Lu wrote:

- Today, Ronic and I locked at the amplifier current of 2 A and obtained ~60% coupling after optimizing the CEP (see Figure 1).

- The injected power is 10 W at 2 A. We measured only 14 kW inside the cavity, which corresponds to an effective gain of 1,400 and a full gain of 2,300. The cavity finesse is 23,000 and the normal gain should be around 6,200.

- We found fluctuations in transmission, possibly because of mode degradation. Tomorrow we will use D-shape mirrors to suppress high-order modes and optimize alignment and locking.

Xinyi Lu wrote:

Today, Ronic and I installed the new telescope and locked the cavity.

- We locked at the amplifier current of 1 A and obtained 32% of coupling. (see Figure 1)

- The telescope was designed for a current of 2 A (output power ~10 W). To inject this power, we need to add some filters to devices.

- For CEP tuning, when we changed the AOM frequency while cavity locking, sometimes it caused unlock and power drops. It will be dangerous in high-power cases. So it's better to optimize the AOM frequency in low power and just tune the laser current in high power. Now the current variation range of the menhir laser is 750mA to 950mA.

Xinyi Lu wrote:

- These days, Ronic, Fatematuj and I measured the beam parameters of the output of the third-stage amplifier.

- We used 2 wedges and reflection filters to reduce the intensity on the CCD.

- We measured multiple points at pump current of 2 A (output power ~10 W). The waist diameter of the output is w_x = 792.26 um, w_y=873.90 um.

- The next step is to design the telescope and improve the coupling efficiency.

 

 

 

 

 

  13   Wed Oct 31 13:42:22 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomPolarization frequency

Check of the frequency of the onefive locked on each polarization of the cavity (tilt a waveplate by 45°).

Frequency repetition rate : 133.335 MHz on spectrum analyzer for both polarization locked.

  14   Wed Oct 31 13:43:03 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomPolarization frequency

Measurement on 30/10/18.

Loïc Amoudry wrote:

Check of the frequency of the onefive locked on each polarization of the cavity (tilt a waveplate by 45°).

Frequency repetition rate : 133.335 MHz on spectrum analyzer for both polarization locked.

 

  2   Mon Oct 1 11:30:39 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomPhase measurement with HASO

2 measurement.

First one locked, in transmission of M2 with 2nd stage 0A.Total 89 cm from the big waist (planar mirrors). 2 wedges used.

Second one at input beam with the same power. Datas taken at the same equivalent position than the first one. 3 wedges used.

  3   Mon Oct 1 14:27:35 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomPhase measurement with HASO

* Loic has to fix the number of files (3) regarding the number of measurements (2)

* splitted Intensity and phase HASO files

* image for each file

Loïc Amoudry wrote:

2 measurement.

First one locked, in transmission of M2 with 2nd stage 0A.Total 89 cm from the big waist (planar mirrors). 2 wedges used.

Second one at input beam with the same power. Datas taken at the same equivalent position than the first one. 3 wedges used.

 

  4   Mon Oct 1 15:19:48 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomPhase measurement with HASO

with just intensity, the coupling is 98%, and with phase, x direction coupling is 95%, y direction 97%, so the telescope is good.

Loïc Amoudry wrote:

* Loic has to fix the number of files (3) regarding the number of measurements (2)

* splitted Intensity and phase HASO files

* image for each file

Loïc Amoudry wrote:

2 measurement.

First one locked, in transmission of M2 with 2nd stage 0A.Total 89 cm from the big waist (planar mirrors). 2 wedges used.

Second one at input beam with the same power. Datas taken at the same equivalent position than the first one. 3 wedges used.

 

 

  63   Thu Jun 10 17:55:15 2021 Ronic ChicheFixedinfolasers and opticsOptical roomOrigami SN2440 133MHz laser tests

Today, we received the Origami SN2440 133MHz oscillator from repair (NKT mentioned a defective wire inside the controller....)

We immediately installed it behind an isolator (Faraday rotator+PBS+halfwave plate).

the output power is around 60mW.
the spectrum is around 1030nm
the repetition rate is around 133.33MHz

  62   Tue Jan 12 10:44:57 2021 Ronic ChicheFixedissuelasers and opticsOptical roomOrigami SN2440 133MHz laser power measurment... and failure

we received from NKT the Origami SN2440 133MHz oscillator from repair on 2020/12/17
and we measured the power trend from 2020/12/21 until the whole winter Hollyday, during 14 days and 17h (1 273 334 s).
the power was quite constant, about 58mW (the measurement has been done with the OD2 filter and using a (x20) factor in the software to compensate for it),
except for some peaks at the beginning of the measurement to 60.7mW.
I tried to reproduce these peaks by switching on several lights in the room and the airflow ceil but the effect is neglectable,
thus these peaks seem to really come from the oscillator power.

at the beginning of January, the oscillator suffered a back reflection from the rugged anodized convex surface of the power meter (apparently not from the OD2 filter itself) and the laser stopped immediately to modelock.
the laser power dropped to 0 and the laser controller led started to blink red. NKT has been contacted and they think it could be the laser pump diodes that have been damaged.

for the future, we will have to install a fixed optical isolator from the start, even for short operations.

 

  39   Tue Dec 17 12:16:54 2019 Ronic ChicheFixedinfolasers and opticsOptical roomOptical amplifier polarization state

We placed a PBS + 2 photodiodes (PhD1, PhD2) at the output of the amplifier to check how the polarization of the amplifier changes with power.

example with 2nd stage @ 6A :
PhD1 = 24.7 mV
PhD2 = 8.9 mV
PhD1/PhD2 = 2.78

and with 3rd stage @ 2A :
PhD1 = 353 mV
PhD2 = 82.8 mV
PhD1/PhD2 = 4.26

Conclusion : we must adapt the quarter and half waveplates for each input power to be always matched with cavity polarization !!!
One could also study how the amplifier polarization changes during time and temperature.

  40   Tue Dec 17 17:54:38 2019 Ronic ChicheFixedinfolasers and opticsOptical roomOptical amplifier polarization state

The polarimeter was giving a strange 50% of DOP of the light coming from the cavity.
we had to calibrate (LONG calibration process with care) the polarimeter to get a proper 100% of DOP !
the polarimeter needs also a good alignment with 2 mirrors, a colimated beam and a max power on photodiode between 0.7 and 0.8 (use electronic gain to adapt the level)

at low power (1.5kW inside cavity), the cavity is almost vertically polarized (89°).

Ronic Chiche wrote:

We placed a PBS + 2 photodiodes (PhD1, PhD2) at the output of the amplifier to check how the polarization of the amplifier changes with power.

example with 2nd stage @ 6A :
PhD1 = 24.7 mV
PhD2 = 8.9 mV
PhD1/PhD2 = 2.78

and with 3rd stage @ 2A :
PhD1 = 353 mV
PhD2 = 82.8 mV
PhD1/PhD2 = 4.26

Conclusion : we must adapt the quarter and half waveplates for each input power to be always matched with cavity polarization !!!
One could also study how the amplifier polarization changes during time and temperature.

 

  42   Thu Dec 19 09:21:45 2019 Ronic ChicheFixedinfolasers and opticsOptical roomOptical amplifier polarization state

polarization state of the cavity at higher power : 20kW, 30kW  and 33kW (slight CEP and alignment optimization) :
the polarization state changes only a little to ~ 87° and is almost linear.

Ronic Chiche wrote:

The polarimeter was giving a strange 50% of DOP of the light coming from the cavity.
we had to calibrate (LONG calibration process with care) the polarimeter to get a proper 100% of DOP !
the polarimeter needs also a good alignment with 2 mirrors, a colimated beam and a max power on photodiode between 0.7 and 0.8 (use electronic gain to adapt the level)

at low power (1.5kW inside cavity), the cavity is almost vertically polarized (89°).

Ronic Chiche wrote:

We placed a PBS + 2 photodiodes (PhD1, PhD2) at the output of the amplifier to check how the polarization of the amplifier changes with power.

example with 2nd stage @ 6A :
PhD1 = 24.7 mV
PhD2 = 8.9 mV
PhD1/PhD2 = 2.78

and with 3rd stage @ 2A :
PhD1 = 353 mV
PhD2 = 82.8 mV
PhD1/PhD2 = 4.26

Conclusion : we must adapt the quarter and half waveplates for each input power to be always matched with cavity polarization !!!
One could also study how the amplifier polarization changes during time and temperature.

 

 

ELOG V3.1.4-395e101