    Reply   Damage on mirror surface , posted by Manar Amer at Optical room about lasers and optics 
ThomX injection mirror has been cleaned and placed again inside the optical cavity.

This time to avoid the damaged spot I have displaced the mirror mount horizontally to have a distance between center of the beam and the spot
~ 2.5 - 3 mm.
    Reply   M1 ThomX used while shifted from damaged spot , posted by Manar Amer at Optical room about lasers and optics Screenshot_2022-09-19_2_172943.pngScreenshot_2022-09-19_1_171834.png00mode_LOCK.jpg00mode_LOCK_saturaed.jpg
Yesterday , we locked the cavity and we see a sign of a high finesse on the transmission signal, but no measurement of Finesse was done.

we have a coupling of ~ 45%, which is a loss of 20% from the previous coupling of 60%

an estimate done by Ronic MATLAB simulation for the coupling drop where we have 200 pp additional losses and gain of 2.6 k we should get a transmission
    Reply   M1 ThomX used while shifted from damaged spot , posted by Manar Amer at Optical room about lasers and optics 
To compare between the 2 images of the cavity mode:


the mode by itself has an integration time of 0.06 ms, position (x, y) = (1142.969, -53.932) um on the beam profiler 
the
mode saturated with the spot next to it almost at max intensity has an integration time of 50 ms, position (-3700, -2000) um


comparing
    Reply   M1 ThomX used while shifted from damaged spot , posted by Manar Amer at Optical room about lasers and optics 00mode_damaged_0_26ms.jpg00mode_damaged_saturted_50ms.jpgreflection_noLock.jpgreflection_Lock_0_15ms.jpg
On Wednesday 21st , I opened the cavity did an additional 2 mm shift of the injection mirror and put it under vacuum again.

Locked the cavity, and observed the transmitted beam.

The second spot is still visible on the beam profiler , the distance difference between the 2 spots is ~ 5.2 mm (the same as before )
    Reply   M1 Gamma Factory, posted by Manar Amer at Optical room about lasers and optics 00mode_ThomX_Mirror_degeneracySpot_1point04ms.jpg00mode_ThomX_Mirror_degeneracySpot_saturateat50.jpgCoupling_mirror_.jpeg
In the morning, Vacuum broken and rotated M1 ThomX 90 degrees clockwise, locked the cavity in air and we observe a degeneracy close to the fundamental
mode.

In the afternoon, I cleaned M1 from Gamma factory using pure ethanol and pure water with the spincoater
    Reply   M2 ThomX spherical Cleaning , posted by Manar Amer at Optical room about lasers and optics 6x
Today, M2 Spherical-3 from ThomX that was installed inside the SBox was removed, there was one big dust on the surface of the mirror, mirror was cleaned

using pure ethanol and pure water with spincoater. (images taken with arrow far from us)

M1 from Gamma factory, fixed with the addition of the ring with 3 screws.
Entry   Cavity polarization states (Koheras), Finesse of 2 polarization states, posted by Loïc Amoudry at Optical room about lasers and optics Finesse_higher.isfFinesse_lower.isfThe2polarizations_states_direct_trans.isfThe2polarizations_states_direct_trans.png
Measurments of Finesse with the 2 polarization states, let's call them H (higher) and L (lower): 24500 for the H and 23500 for the L.

We checked the polarization states in transmission of the FP cavity after a PBS. The H was stronger in PBS trans and the L stronger in PBS ref.

We measured the power in reflection of the PBS and added a half WP that we aligned with the PBS polarization. Then, to get the maximum power
Entry   Phase measurement with HASO, posted by Loïc Amoudry at Optical room about lasers and optics 180928_M2_0A.has1801001_input_0A.has180928_M2_trans_0A.txt
2 measurement.

First one locked, in transmission of M2 with 2nd stage 0A.Total 89 cm from the big waist (planar mirrors). 2 wedges used.

Second one at input beam with the same power. Datas taken at the same equivalent position than the first one. 3 wedges used.
    Reply   Phase measurement with HASO, posted by Loïc Amoudry at Optical room about lasers and optics 8x
* Loic has to fix the number of files (3) regarding the number of measurements (2)

* splitted Intensity and phase HASO files

* image for each file
    Reply   Phase measurement with HASO, posted by Loïc Amoudry at Optical room about lasers and optics 
with just intensity, the coupling is 98%, and with phase, x direction coupling is 95%, y direction 97%, so the telescope is good.




Loïc
Amoudry wrote:



* Loic has to fix the number of files (3) regarding the number of
Entry   Polarization optimization, posted by Loïc Amoudry at Optical room about lasers and optics 
Optimization of the polarization has been made the 03/10/18. Checked in reflection of the cavity in reflection&transmission of a PBS, locked and
unlocked. Only with 2nd stage.

Ratio values are reflection of PBS divided by transmission or the opposite.
Entry   Record power-up, posted by Loïc Amoudry at Optical room about lasers and optics 
Measurement made on 03/10/18 (nothing has been done since there).

Stable power in the cavity of 225 kW.




3rd stage current
Entry   Motors for D-shaped mirrors, posted by Loïc Amoudry at Optical room about mechanics | lasers and optics tek00000.pngtek0000CH1.isftek0000CH2.isftek0000CH3.isftek0000CH4.isf
Motors have been installed on 16/10/18. No problem with them.

Effect of the motors tested on 17/10/18. No improvement. But they give the possibility to perfectly cut HOM or let them go through as show
the following picture of a 2.2 mode at ~340 mW in trans and 70% coupling @4A.
Entry   Transmission vs D-shape position at different powe, posted by Loïc Amoudry at Optical room about lasers and optics Plot_matlab.JPGmotors.xlsxPower_stored_vs_Dshape_mirrors_position.m
Measurements have been done on 18/10/18.

Datas are on excel file, also matlab file.
Entry   Beam size vs D-shape position, posted by Loïc Amoudry at Optical room about lasers and optics 181018_4A_no_cut.PNG
Measurement done on 18/10/18.

At high power, the shape of the 0.0 mode does not change. The D-shape only generate losses in the cavity. Then the power stored in the cavity
decrease. As with this configuration, the cavity beam size decrease when power increase, the beam size decreased.
    Reply   Beam size vs D-shape position, posted by Loïc Amoudry at Optical room about lasers and optics size_vs_position_and_power.JPG
Matlab code for size vs position and power : 

clear all
close all
Entry   Power measurement with D-shape, posted by Loïc Amoudry at Optical room about mechanics | lasers and optics 
Measurements of lot of points with D-shape mirrors well positionned.

Power not optimized to the best but almost. (@4A could have 350 mW).




I (A)
Ptrans (mW)
 Coupling
Entry   Finesse vs power by difference between main and second resonance, posted by Loïc Amoudry at Optical room about lasers and optics CrossSecondaryResonance.mGetCavity.m
Measurements show that ratio decrease versus power. BUT, the second resonance measurement induce lower power in the cavity so the ratio is not directly
true.

Also, simulation of the main/second resonance power by Pierre's simulation has shown ratio ~50, ~47.6 and 43.5 respectively for 0A, 2A and
Entry   Polarization frequency, posted by Loïc Amoudry at Optical room about lasers and optics 
Check of the frequency of the onefive locked on each polarization of the cavity (tilt a waveplate by 45°).

Frequency repetition rate : 133.335 MHz on spectrum analyzer for both polarization locked.
    Reply   Polarization frequency, posted by Loïc Amoudry at Optical room about lasers and optics 
Measurement on 30/10/18.




Loïc
Amoudry wrote:



Check of the frequency of the onefive locked on each polarization