Message ID: 61     Entry time: Tue Jan 5 12:38:04 2021
 Author: Manar Amer 
 Status: Fixed 
 Type: report 
 Category: lasers and optics 
 Location: Optical room 
 Title: Transmission Measurements for the Mirrors.  
  • Thursday 10 December 2020:   First Test Measurement - Inside Box
    •  The transmission ( T ) for the mirrors of SBOX and ThomX was measured over a range from the center of the mirror up to 5mm.
    •  
    • The first setup was initially done while the mirrorr is installed inside the metal box, when measuring T we observe that there is an erregular  change in the power measured as we move azay from the center of the mirror, this was due to reflection of light inside the box. The reflection from the mirror was reflected on the metal inside the box and it was measured by the power meter.

               (so for this setup we only took measrements for T along the center of the mirror)

The file stating the measurements is : Test Cavity _ Mirrors transmission _ Inside box

=======================================================================

  • Tuesday 15 December 2020 :  Second Test - Outside Box
    • The setup for T test was moved outside the metal box with the same distances between the injection beam and mirror ~ 50 cm. and the distance between the injection laser and the mirror next to it ~ 9 cm.
    •  
    • Observation for T is done at full angle and at half angle of reflection ( the mirror was tilted to achive this )
    •  
    • we observed that after each change in the distance from the center the power meter needed to be aligned to have the max power.
    • after doing all the measurements for the range from the center we made it again for the center only and the consistance of the transmission was good.

The file stating the measurements is : Test Cavity _ Mirrors transmission _ outside box

note: the laser used for the test is the Koheras

details on the pwer used, mirrors, power meter, filters, ....etc are shown in the Excel data sheets

Done By : Loic , Manar and Ahmed

Attachment 1: Test_Cavity___Mirrors_transmission___Inside_box.xlsx  10 kB
Attachment 2: Test_Cavity___Mirrors_transmission___outside_box.xlsx  12 kB
Attachment 3: set_up_for_transmission_measurement_outside_box.jpeg  182 kB  | Show | Hide all | Show all