  9   Thu Apr 29 06:14:09 2021 Manar AmerFixedreportlasers and opticsOtherMode observed and cavity Lock ** Following up from yesterday an observation
about the TM00 mode : it has been seen with
similar dimensions and its position is the
 20212804_Fundamental_mode_in_cavity_2D_-_Position.png20212804_Cavity_Coupling_with_Koheras.jpg20212804_Temperature_Curve.jpg 
  14   Wed Jun 30 20:29:17 2021 Manar AmerFixedreportlasers and opticsOtherMode observed and cavity Lock New configuration is made. The new mode
observed in NF_Trans is the following :

 NF_Trans__2021-06-25__16-38-56.bmp 
  3   Thu Apr 15 19:45:08 2021 Manar AmerFixedreportmechanicsOtherMirrors Mottors Connections * The connections to control the motors
of the mirrors are connected in the order
stated :
 20210804_Motor_connections_with_mirrors_inside_cavity_box.jpg20210804_Mottor_cables_behind_cavity_box.jpg 
  4   Thu Apr 15 19:54:19 2021 Manar AmerFixedreportmechanics | cablingOtherMirrors Mottors Connections Added the cabling to the categories 




  
  68   Thu Feb 2 16:21:05 2023 BlancFixedinfolasers and opticsOptical roomMesures de Polarisation Lockline et TangorVérification des polarisations après
le combiner pour le Tangor et la Lockline.

On remarque certains paramètres
 11x 
  66   Mon Jan 30 14:39:52 2023 BlancFixedinfolasers and opticsOptical roomMesures de PolarisationEn récupérant la transmission
du miroir M3 sphérique sortant de
la cavité, on a pu faire une mesure
 Paramètres_de_Stokes.pngDOP_DOLP_DOCP.pngPhase_Ellipticité_Êta.pngEllipse_polarisation_cavité.png 
  67   Thu Feb 2 16:06:54 2023 BlancFixedinfolasers and opticsOptical roomMesures de PolarisationCorrection su l'orientation de l'ellipse.




 Ellipse_Polarization_CAVITY.png 
  74   Fri Feb 17 18:02:11 2023 BlancFixedinfolasers and opticsOptical roomMesure polarisation voie de réflexionJ'ai fait les mesures de polarisation
sur la voie de réflexion pour voir
la polarisation avant le glan et après
 18x 
  62   Fri Jan 20 18:01:22 2023 BlancFixedinfolasers and opticsOptical roomMesure de finesseFréquence centrale : 879.48 MHz

FWHM mesurée à la
main sans sweep : 120 / 130 KHz
  
  63   Wed Jan 25 12:12:36 2023 BlancFixedinfolasers and opticsOptical roomMesure de finesse 




 Mesure_de_finesse.png 
  52   Fri Dec 17 17:03:34 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 17 12 2021 reprise réf nouvelle injection pour polarisationTangor : 4.2W
* Réalignement de l'injection
suite aux modifications :
 Screenshot_2021-12-17_0_164132.pngScreenshot_2021-12-17_1_170202.png 
  49   Thu Dec 16 12:12:10 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 16 12 2021 montée en puissanceMatin: 
* Redémarrage Tangor 4W
    - couplage 15%
 Screenshot_2021-12-16_0_115846.png 
  50   Thu Dec 16 12:13:52 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 16 12 2021 montée en puissance 




 Screenshot_2021-12-16_1_121553.png 
  47   Tue Dec 14 12:59:07 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 14 12 2021 injection optimization (and calibration)Matin :

moteurs :
M1, M2 (plans): 5.110500m
 Screenshot_2021-12-14_2_125459.png 
  48   Thu Dec 16 10:40:55 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 14 12 2021 injection optimization (and calibration)Après-midi :
    * Optimisation
de l'alignement en injection pour avoir
 Screenshot_2021-12-14_3_182645.png 
  46   Fri Dec 10 16:57:56 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 10 12 2021 (optimization injection Burst)Manip 10/12/2021
Après recherche de la résonnance
principale
 NF_Refl__2021-12-10__14-25-08.pngtest__2021-12-10__14-25-08_1920_25gm.png 
  32   Tue Oct 5 18:24:53 2021 Ronic ChicheFixedinfolasers and opticsOptical roomLocking with fast analog loop on AOMin addition to the standard locking scheme
with the GHz laser PZT,

we added an AOM after the PDH modulation
  
  18   Tue Sep 7 17:48:50 2021 BlancFixedinfolasers and optics | detectors and electronicsOptical roomLocking of the cavity and setup changingWe change the setup a bit to have the same
polarization on the photodiode and the PDH
(in reflexion of the cavity). CF: schematics
 scope_0.png 
  10   Fri Apr 30 17:41:10 2021 Aurélien MartensFixedreportlasers and optics | detectors and electronicsOptical roomLock, cavity length and finesse estimateRonic just suceeded in obtaining a reasonably
good lock on the cavity. Air flow is switched
off.
 IMG_20210430_173331.jpg 
  16   Tue Sep 7 11:05:47 2021 Aurélien MartensFixedreportlasers and optics | detectors and electronicsOptical roomLock, cavity length and finesse estimatewe observed that the transmitted light
coming from the cavity is made of the cavity
mode light and the scattered light on the
  