38
|
Wed Oct 20 15:27:51 2021 |
Ronic Chiche | Fixed | report | lasers and optics | Optical room | Amplitude GHz oscillator phase noise with PZT effect |
Thorlabs PZT datasheet.
Reference: PC4QR
| Ronic Chiche wrote: |
|
3 phase noise measurements made on the Amplitude GHz oscillator with different PZT configurations :
- black curve: PZT connector is open
- green curve: PZT connector is shorted by 50 ohms
- blue curve: PZT is excited by 100mVrms of white noise coming from a generator.
on the blue curve, one can clearly see a phase noise increase in the region 10kHz - 1MHz but it is not evident the peaks seen with the PZT open or shorted are related to the peaks excited with the noise injected on the PZT.
with a PZT not excited, one can just observe that the phase noise is decreasing a lot around 10kHz to reach the reference oscillator phase noise floor and then increase again exactly when the PZT resonant frequencies appear, between 20kHz and 200kHz.... reaching at the end the phase noise detection floor.
|
|
| Attachment 1: PC4QR-SpecSheet-1.pdf
|
|
46
|
Fri Dec 10 16:57:56 2021 |
Kevin Dupraz | Under Process | report | lasers and optics | Optical room | Manip 10 12 2021 (optimization injection Burst) |
Manip 10/12/2021
Après recherche de la résonnance principale
- MCS miroirs plan à 5.110250mm (en derive vers les negatifs)
- pic transmission à 5V
- couplage ~7%
Décalage du spot (très faible, à peine visible à la camera NF_refl)
- grosse perte en transmission ==> alignement OK
Grossissement du spot
* début taille_sigma à 0.220mm (calibration NF_Refl)
* nouvelle position en éloignant la lentille 1 taille à 0.300mm ==> max transmission ~3V,
- Walk en vertical ==> pas d'amélioration et dégradation.
- Divergent en entrée de cavité (test à la carte)
* En rapprochant la lentille 1 passage pas le minimum de taille vers 0.190 mm
* Depassement du minimum. Taille faisceau 0.317 mm (0.380 mm sur y) ==> max transmission ~2.5V,
- Walk en vertical ==> pas d'amélioration et dégradation.
- divergence pas apréciable
* On se met au minimum de taille sur la caméra. Taille faisceau 0.178 mm (0.170mm sur y) ==> max transmission ~5.3V
- Couplage 25% max (beaucoup de fluctuations).
- Gros décalage vertical (~ 1 sigma) ne provoque pas une baisse du transmis.
- Décalage horizontal sensible.
Après mangé Ronic a "lockée" la cavité. Le gain de la photodiode en transmission a été changé et la tension sur la résonnance principale est d'environ 600 - 800 mV
deux images ont été faites en entrée de cavité à environ 18 cm du splitter pour la camera de test afin d'être à peu près à une distance équivalente du miroir d'injection M1.
|
| Attachment 1: NF_Refl__2021-12-10__14-25-08.png
|
 |
| Attachment 2: test__2021-12-10__14-25-08_1920_25gm.png
|
 |
|
47
|
Tue Dec 14 12:59:07 2021 |
Kevin Dupraz | Under Process | report | lasers and optics | Optical room | Manip 14 12 2021 injection optimization (and calibration) |
Matin :
moteurs :
M1, M2 (plans): 5.110500m
M3, M4 (sphériques): -2.900000m
- Optimization de la première L/2 (en sortie tangor), le flux est maximisé en sortie du combiner.
* mesure sortie Tangor (Power meter de puissance) : 85 + 90 mW
* mesure sortie Tangor (Power meter PH100-Si-HA-OD1 "226316" + OD2) : 4.11mW
* mesure avant combiner, entre miroirs de replis Newfocus (Power meter + OD2) : 3.53mW
* mesure après pick-off (injection cavité) (Power meter + OD2) : 2.93mW
- Optimization du couplage avec la polarisation :
* modification de la polarisation injecté dans la cavité
* optimisation du flux reçu sur la photodiode en réflection
* mesure du couplage.
** Optimum trouvé à 15% de couplage (en scan 5Hz, 0-10V)
Position des lambdas :
- L/2 en injection : 333°
- L/4 en injection : 48°
- L/2 en réflection : 225°
- L/4 en réflection : 54°
- Déplacement de la photodiode en transmission.
* Réglage de l'iris au niveau de la photodiode : Iris ouvert PV résonance 31 mV, Iris fermée 25 mV
- Lock de la cavité.
* transmission : 2.65V (60dB)
* reflection : 16.5% de couplage
* mesure devant la caméra en transmission (Power meter PH100-Si-HA-OD1 "226316" + OD2) : 0.635 uW
* image (P.J) sauvegardé bin |
| Attachment 1: Screenshot_2021-12-14_2_125459.png
|
 |
|
48
|
Thu Dec 16 10:40:55 2021 |
Kevin Dupraz | Under Process | report | lasers and optics | Optical room | Manip 14 12 2021 injection optimization (and calibration) |
Après-midi :
* Optimisation de l'alignement en injection pour avoir le plus de puissance en transmission
* Optimisation (vérification) de la polarisation pour avoir le plus de couplage.
* tension photoDiode en réflexion (50 Ohm) : max à 136 mV (couplage 20%)
* tension Photodiode transmission PDA100A-EC(50 ohm) : max à 3.6 V (60 dB de gain)
- Passage à 50 W en sortie de TANGOR.
* Vérificaton de l'alignement pour avoir le plus de puissance en transmission.
* Vérification de la polarisation pour avoir le plus de couplage.
* tension Photodiode transmission PDA100A-EC(50 ohm) : max à 4.9 V (40 dB de gain)
* Changement de la polarisation sur la ligne de réflexion (uniquement la lambda/4) pour ne pas saturer la photodiode.
* tension photoDiode en réflexion (50 Ohm) : max à 136 mV (couplage 12%)
* Références caméras :
> NF_refl : x0 = 11.9 mm; y0 = 8.65 mm; sigma_x = 0.185 mm; sigma_y = 0.167 mm.
> FF_refl : x0 = 2.44 mm; y0 = 0.61 mm; sigma_x = 0.241 mm; sigma_y = 0.259 mm.
> NF_inj : x0 = 8.57 mm; y0 = 5.40 mm; sigma_x = 0.123 mm; sigma_y = 0.127 mm.
- Plus de splitter 90/10 en sortie de TANGOR, 4.5 W en sortie TANGOR :
* Compensation du walk-of introduit par ce splitter.
* réalignement de la cavité après locking
* Photodiode Transmission (60dB + densité OD30) : max 2V
* Reflection (près changement de pile) : 18% de couplage (tension max 136 mV, 50 Ohm)
| Kevin Dupraz wrote: |
|
Matin :
moteurs :
M1, M2 (plans): 5.110500m
M3, M4 (sphériques): -2.900000m
- Optimization de la première L/2 (en sortie tangor), le flux est maximisé en sortie du combiner.
* mesure sortie Tangor (Power meter de puissance) : 85 + 90 mW
* mesure sortie Tangor (Power meter PH100-Si-HA-OD1 "226316" + OD2) : 4.11mW
* mesure avant combiner, entre miroirs de replis Newfocus (Power meter + OD2) : 3.53mW
* mesure après pick-off (injection cavité) (Power meter + OD2) : 2.93mW
- Optimization du couplage avec la polarisation :
* modification de la polarisation injecté dans la cavité
* optimisation du flux reçu sur la photodiode en réflection
* mesure du couplage.
** Optimum trouvé à 15% de couplage (en scan 5Hz, 0-10V)
Position des lambdas :
- L/2 en injection : 333°
- L/4 en injection : 48°
- L/2 en réflection : 225°
- L/4 en réflection : 54°
- Déplacement de la photodiode en transmission.
* Réglage de l'iris au niveau de la photodiode : Iris ouvert PV résonance 31 mV, Iris fermée 25 mV
- Lock de la cavité.
* transmission : 2.65V (60dB)
* reflection : 16.5% de couplage
* mesure devant la caméra en transmission (Power meter PH100-Si-HA-OD1 "226316" + OD2) : 0.635 uW
* image (P.J) sauvegardé bin
|
|
| Attachment 1: Screenshot_2021-12-14_3_182645.png
|
 |
|
49
|
Thu Dec 16 12:12:10 2021 |
Kevin Dupraz | Under Process | report | lasers and optics | Optical room | Manip 16 12 2021 montée en puissance |
Matin:
* Redémarrage Tangor 4W
- couplage 15%
- max photodiode Transmission (60dB, OD30) : 1.9V
* Montée en Puissance 20 W
- début des effets thermiques : dérive thermique (la cavité chauffe et le piezo doit suivre, la tension baisse)
- max photodiode Transmission (40dB, OD30) : 1.6 V
- Reflection (~12%) ~140mV max:
* quand résonance => plus de signal sur la reflexion que hors lock.
- Hypothèses:
saturation de la photodiode ==> avec une densité OD10 (même effet)
effet de polarisation: essaie avec la densité OD10 (absorbante) dans le faisceau au niveau de la lambda/4 en reflection ==> la densité est brulée (ne tient pas la puissance)
==> essayer avec densité réflective.
* impossible de locker sur le max de transmission |
| Attachment 1: Screenshot_2021-12-16_0_115846.png
|
 |
|
50
|
Thu Dec 16 12:13:52 2021 |
Kevin Dupraz | Under Process | report | lasers and optics | Optical room | Manip 16 12 2021 montée en puissance |
| Kevin Dupraz wrote: |
|
Matin:
* Redémarrage Tangor 4W
- couplage 15%
- max photodiode Transmission (60dB, OD30) : 1.9V
* Montée en Puissance 20 W
- début des effets thermiques : dérive thermique (la cavité chauffe et le piezo doit suivre, la tension baisse)
- max photodiode Transmission (40dB, OD30) : 1.6 V
- Reflection (~12%) ~140mV max:
* quand résonance => plus de signal sur la reflexion que hors lock.
- Hypothèses:
saturation de la photodiode ==> avec une densité OD10 (même effet)
effet de polarisation: essaie avec la densité OD10 (absorbante) dans le faisceau au niveau de la lambda/4 en reflection ==> la densité est brulée (ne tient pas la puissance)
==> essayer avec densité réflective.
* impossible de locker sur le max de transmission
|
|
| Attachment 1: Screenshot_2021-12-16_1_121553.png
|
 |
|
51
|
Thu Dec 16 18:16:12 2021 |
Blanc | Under Process | info | lasers and optics | Optical room | Changements et ajout d'optiques |
Optiques ajoutées :
- splitter 90/10 au début de la voie de réflexion, pour éjecter le flux et éviter que les lambdas ne chauffent.
- deuxième polariser pour avoir une meilleur extinction.
- splitter 10/90 après la première lentille sur la voie d'injection qui envoie 10% du flux sur la photodiode de réflexion.
Optiques changées :
- miroir HR à la place d'un splitter 97/3 à 0° que l'on utilisait à 45° dont le faisceau en transmission allait sur la photodiode de réflexion.
- splitter 70/30 à la place d'un splitter 98/2 à 0° que l'on utilisait à 45° qui divisait champ proche/champ lointain.
Mise en place d'un dump pour le faisceau réfléchi par le splitter 90/10 du début de la voie de transmission. |
|
52
|
Fri Dec 17 17:03:34 2021 |
Kevin Dupraz | Under Process | report | lasers and optics | Optical room | Manip 17 12 2021 reprise réf nouvelle injection pour polarisation |
Tangor : 4.2W
* Réalignement de l'injection suite aux modifications :
- Changement de l'ordre des waveplate maintenant : lambda/4, puis lambda/2
- suppression de l'iris au niveau de la transmission, perte d'au moins un facteur sur le signal en transmission
- reprise des références sur les caméras en reflection. (NF:[10.2548, 6.1092], FF:[1.9881, 1.5929])
* Optimisation du couplage en pointé, puis en polarisation
--> Maximum de reflection dans la photodiode (transmission maximale)
- couplage mesuré en réflection : 32% +- 4%(PV) max (180 mV)
- max intensité en transmission (0 dB, sans densité) : 1.1V
Tangor : 20W
- Montée en puissance: grosse dérive thermique mais le signal en réflection est "correcte".
- transmission (30dB, OD30) : pic max à 2.6 V
"puissance stable", 1.88V
- couplage : 25% (40mV/160mV)
On part en vacances ^^ |
| Attachment 1: Screenshot_2021-12-17_0_164132.png
|
 |
| Attachment 2: Screenshot_2021-12-17_1_170202.png
|
 |
|
55
|
Thu May 19 10:04:05 2022 |
Blanc | Under Process | report | lasers and optics | Optical room | Changement cadence et duty cycle + ref cam |
- Prise des nouvelles ref caméras : NF{709;371} , FF{389;280} [Image 1 & 2]
- Placement d'une photodiode rapide en transmission pour voir le train d'impulsions du Tangor à travers la cavité.
- Lock du Tangor en quasi-continue, P_moy = 2W, 100 kHz.
- Baisse de la cadence à 10 kHz, duty cycle 50% (sur le soft), 25 % (sur l'oscillo), à comprendre, P_moy = 1W [Image 3,4,5]
- On observe que la forme du burst a un impact sur le PDH et par consequent sur le piezo [Image3 & 4]
- Possibilité d'ajuster cela en jouant sur le gain du PID, mais si on baisse en cadence, en duty cycle et qu'on monte en puissance cela suffira ? |
| Attachment 1: Ref_NF_inj_18_05_2022.png
|
 |
| Attachment 2: Ref_FF_inj_18_05_2022.png
|
 |
| Attachment 3: Screenshot_2022-05-18_2_165654.png
|
 |
| Attachment 4: Screenshot_2022-05-18_5_173440.png
|
 |
| Attachment 5: Screenshot_2022-05-18_6_174947.png
|
 |
|
56
|
Tue Jun 14 11:16:37 2022 |
Blanc | Under Process | issue | lasers and optics | Optical room | Problèmes Cavitée |
Précédentes ref pour lesquelles la manip fonctionnait :
Caméras :
709 ; 371 NF
389 ; 280 FF
Moteurs :
5.091422 ; 5.091422 ; -2.9 ; -2.9
Problèmes dans la cavité :
- Pas de beating de mode alors que l'on est sur les références d'injection
- Faisceau carré en transmission
- Les softs étaient ouverts et la clé de la lockline tournée en position laser alors que j'avais tout fermé à la fin des manips et mis en standby...
|
| Attachment 1: Transmission.PNG
|
 |
| Attachment 2: NF_inj.PNG
|
 |
| Attachment 3: FF_inj.PNG
|
 |
|
57
|
Tue Jan 10 18:04:20 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Données pour la baisse en cadence |
Position Cam :
692;363 NF_inj
520;193 FF_inj
Position Moteurs :
5.085435
5.085435
-2.9
-2.9
Spectre de l'oscillateur ~6 nm largeur (correct, cf image)
Durée d'impulsion : 3.35 ps (cf image)
On peut également voir que la forme du mode TEM00 a changé (cf image sur la caméra de transmission), dû à un endommagement d'un/des miroirs de la cavité. On voit également que rien n'est visible sur la caméra de réflection.
AOM 100 Hz , t_burst = 5 µs, PP 50 kHz : P_burst = 0.15 W, P_burst_PhD = 0.47 mW
AOM 100 Hz , t_burst = 5 µs, PP 50 kHz : P_burst = 0.25 W, P_burst_PhD = 0.75 mW
AOM 100 Hz , t_burst = 5 µs, PP 50 kHz : P_burst = 0.8 W, P_burst_PhD = 2.09 mW
J'ai également mis les mesures de long-term, ainsi que les burst correspondants. |
| Attachment 1: Spectre_oscillateur_GHz_10_01_2023.PNG
|
 |
| Attachment 2: Pulse_Width_Compressor_10000µm_10_01_2023.png
|
 |
| Attachment 3: Mode_TEM00_lockline_lockée_après_cassage_cavité_trans2_10_01_2023.PNG
|
 |
| Attachment 4: Mode_TEM00_lockline_lockée_après_cassage_cavité_trans_10_01_2023.PNG
|
 |
| Attachment 5: Mode_TEM00_lockline_lockée_après_cassage_cavité_NF_refl_10_01_2023.PNG
|
 |
| Attachment 6: Long-term_Histo.png
|
 |
| Attachment 7: Tangor_and_Cavity_burst.png
|
 |
|
58
|
Wed Jan 18 17:52:33 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Tests et mesures lockline |
Mesures de puissance fibres lockline :
AVANT POLISSAGE
injection premier étage : 15 mW
sortie premier étage : 63 mW
Sortie AOM à 250 MHz : 2.5 mW
Sortie EOM : 1.3 mW
Sortie : 20 =>40 mW
APRES POLISSAGE
injection premier étage : 15 mW
sortie premier étage : 63 mW
Sortie AOM à 159 MHz : 2.9 mW Sortie AOM à 250 MHz : 15 mW
Sortie EOM : 1.5 mW Sortie EOM : ?
Sortie : 47 mW Sortie : 54 mW
Même après polissage de la fibre de sortie du premier étage, on observe que la puissance diminue d'un facteur 4 à travers l'AOM drivé à sa fréquence nominale. |
|
59
|
Thu Jan 19 14:56:27 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Tests et mesures lockline |
| Blanc wrote: |
|
Mesures de puissance fibres lockline :
AVANT POLISSAGE
injection premier étage : 15 mW
sortie premier étage : 63 mW
Sortie AOM à 250 MHz : 2.5 mW
Sortie EOM : 1.3 mW
Sortie : 20 =>40 mW
APRES POLISSAGE
injection premier étage : 15 mW
sortie premier étage : 63 mW
Sortie AOM à 159 MHz : 2.9 mW Sortie AOM à 250 MHz : 15 mW
Sortie EOM : 1.5 mW Sortie EOM : ?
Sortie : 47 mW Sortie : 54 mW
Même après polissage de la fibre de sortie du premier étage, on observe que la puissance diminue d'un facteur 4 à travers l'AOM drivé à sa fréquence nominale.
|
|
| Attachment 1: AOM_250_drivé_à_160_MHz_sortie_fibre_bleue.png
|
 |
| Attachment 2: schéma_mesures_lockline_18_01_2023.PNG
|
 |
|
60
|
Thu Jan 19 15:05:58 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Suite Mesures de puissance lockline |
Changement d'AOM pour passer sur l'AOM 200 MHz.
Le schéma suivant récapitule les mesures prises.
Il y a eu une chute de la puissance en sortie de la fibre bleue de la lockline.
1) à la vue des mesures, il semble que le deuxieme etage de la lockline soit bien fatigué
2) La fibre bleue s'est déteriorée d'un seul coup...
Par contre il n'y a plus les modulations observées précédement sur la photodiode grâce au changement d'AOM, à la fois à 159 MHz et 200 MHz.
|
| Attachment 1: schéma_mesures_lockline_19_01_2023.PNG
|
 |
| Attachment 2: 20230119_124318.jpg
|
 |
|
61
|
Thu Jan 19 18:20:02 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Suite Mesures de puissance lockline |
| Blanc wrote: |
|
Changement d'AOM pour passer sur l'AOM 200 MHz.
Le schéma suivant récapitule les mesures prises.
Il y a eu une chute de la puissance en sortie de la fibre bleue de la lockline.
1) à la vue des mesures, il semble que le deuxieme etage de la lockline soit bien fatigué
2) La fibre bleue s'est déteriorée d'un seul coup...
Par contre il n'y a plus les modulations observées précédement sur la photodiode grâce au changement d'AOM, à la fois à 159 MHz et 200 MHz.
|
Repolissage de la fibre bleue et de la fibre de sortie du deuxieme étage.
Fin des manips :
sortie deuxieme étage : 60 mW
sortie fibre bleue : entre 50 et 55 mW |
|
62
|
Fri Jan 20 18:01:22 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Mesure de finesse |
Fréquence centrale : 879.48 MHz
FWHM mesurée à la main sans sweep : 120 / 130 KHz
=> Finesse = FSR/FWHM = ~7000
Apporter un analyseur de spectre pour régler le sweep du Marconi pour une mesure de finesse plus précise. |
|
63
|
Wed Jan 25 12:12:36 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Mesure de finesse |
| Blanc wrote: |
|
Fréquence centrale : 879.48 MHz
FWHM mesurée à la main sans sweep : 120 / 130 KHz
=> Finesse = FSR/FWHM = ~7000
Apporter un analyseur de spectre pour régler le sweep du Marconi pour une mesure de finesse plus précise.
|
On a pu mesurer la FWHM à l'aide du sweep : 124kHz
On obtient une finesse de 7300 en ayant traité les données. |
| Attachment 1: Mesure_de_finesse.png
|
 |
|
64
|
Wed Jan 25 12:18:25 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Position des miroirs et gain photodiode |
Nouvelle position des miroirs :
5.198865 ; 5.199305 ; -2.9 ; -2.9
Photodiode de transmission, gain de 60, 1MOhm osc : 2.5 V
Photodiode de transmission, gain de 50, 1MOhm osc : 800 mV |
|
65
|
Thu Jan 26 22:49:46 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Nouvelles positions moteurs et calibrations camera |
Positions moteurs :
5.198865 ; 5.198596 ; -2.9 ; -2.9
Calibration caméras :
NF_inj : 957;616
FF_inj : 328;2790
On a réussi à avoir un signal d'erreur très proche de la théorie, ainsi qu'un lock très propre. (Cf images)
En regardant laTF de la lockline, on a vu la modulation à +ou- 159 MHz autour du pic à 880 MHz de l'oscillateur GHz, mais cela n'explique pas l'offset sur le PDH.
Hypohèse à vérifier : le câble d'alimentation du PDH qui est défectueux. |
| Attachment 1: Lock_lockline.png
|
 |
| Attachment 2: Signal_d'erreur_PDH.png
|
 |
|
66
|
Mon Jan 30 14:39:52 2023 |
Blanc | Fixed | info | lasers and optics | Optical room | Mesures de Polarisation |
En récupérant la transmission du miroir M3 sphérique sortant de la cavité, on a pu faire une mesure de polarisation de la cavité. (Cf courbes ci-dessous)
La moyenne pour chaque courbe donne :
S0 = 1.0058 ; S1 = 0.4643 ; S2 = 0.0488 ; S3 = 0.8913
DOP = 1.0062 ; DOLP = 0.4669 ; DOCP = 0.8913
Phase = 86.8705° ; Ellipticité = 0.6076 ; Êta = 2.9962
Température de la tête du polarimètre : 39.9815
Par reconstruction, on obtient l'ellipse donnée en bleu ci-dessous. |
| Attachment 1: Paramètres_de_Stokes.png
|
 |
| Attachment 2: DOP_DOLP_DOCP.png
|
 |
| Attachment 3: Phase_Ellipticité_Êta.png
|
 |
| Attachment 4: Ellipse_polarisation_cavité.png
|
 |