  38   Wed Oct 20 15:27:51 2021 Ronic ChicheFixedreportlasers and opticsOptical roomAmplitude GHz oscillator phase noise with PZT effect

Thorlabs PZT datasheet.
Reference: PC4QR

Ronic Chiche wrote:

3 phase noise measurements made on the Amplitude GHz oscillator with different PZT configurations :

- black curve: PZT connector is open
- green curve: PZT connector is shorted by 50 ohms
- blue curve: PZT is excited by 100mVrms of white noise coming from a generator.

on the blue curve, one can clearly see a phase noise increase in the region 10kHz - 1MHz but it is not evident the peaks seen with the PZT open or shorted are related to the peaks excited with the noise injected on the PZT.

with a PZT not excited, one can just observe that the phase noise is decreasing a lot around 10kHz to reach the reference oscillator phase noise floor and then increase again exactly when the PZT resonant frequencies appear, between 20kHz and 200kHz.... reaching at the end the phase noise detection floor.

 

 

Attachment 1: PC4QR-SpecSheet-1.pdf
  46   Fri Dec 10 16:57:56 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 10 12 2021 (optimization injection Burst)

Manip 10/12/2021
Après recherche de la résonnance principale
    - MCS miroirs plan à 5.110250mm (en derive vers les negatifs)
    - pic transmission à 5V
    - couplage ~7%

Décalage du spot (très faible, à peine visible à la camera NF_refl)
    - grosse perte en transmission ==> alignement OK

Grossissement du spot
    * début taille_sigma à 0.220mm (calibration NF_Refl)
    * nouvelle position en éloignant la lentille 1 taille à 0.300mm ==> max transmission ~3V, 
        - Walk en vertical ==> pas d'amélioration et dégradation.
        - Divergent en entrée de cavité (test à la carte)
    * En rapprochant la lentille 1 passage pas le minimum de taille vers 0.190 mm
    * Depassement du minimum. Taille faisceau 0.317 mm (0.380 mm sur y) ==> max transmission ~2.5V,
        - Walk en vertical ==> pas d'amélioration et dégradation.
        - divergence pas apréciable
    * On se met au minimum de taille sur la caméra. Taille faisceau 0.178 mm (0.170mm sur y) ==> max transmission ~5.3V
        - Couplage 25% max (beaucoup de fluctuations).
        - Gros décalage vertical (~ 1 sigma) ne provoque pas une baisse du transmis.
        - Décalage horizontal sensible.
       

Après mangé Ronic a "lockée" la cavité. Le gain de la photodiode en transmission a été changé et la tension sur la résonnance principale est d'environ 600 - 800 mV

deux images ont été faites en entrée de cavité à environ 18 cm du splitter pour la camera de test afin d'être à peu près à une distance équivalente du miroir d'injection M1.

 

Attachment 1: NF_Refl__2021-12-10__14-25-08.png
NF_Refl__2021-12-10__14-25-08.png
Attachment 2: test__2021-12-10__14-25-08_1920_25gm.png
test__2021-12-10__14-25-08_1920_25gm.png
  47   Tue Dec 14 12:59:07 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 14 12 2021 injection optimization (and calibration)

Matin :

moteurs :
M1, M2 (plans): 5.110500m
M3, M4 (sphériques): -2.900000m


- Optimization de la première L/2 (en sortie tangor), le flux est maximisé en sortie du combiner.
    * mesure sortie Tangor (Power meter de puissance) : 85 + 90 mW
    * mesure sortie Tangor (Power meter PH100-Si-HA-OD1 "226316" + OD2) : 4.11mW
    * mesure avant combiner, entre miroirs de replis Newfocus (Power meter + OD2) : 3.53mW
    * mesure après pick-off (injection cavité) (Power meter + OD2) : 2.93mW

- Optimization du couplage avec la polarisation :
    * modification de la polarisation injecté dans la cavité
    * optimisation du flux reçu sur la photodiode en réflection
    * mesure du couplage.
    ** Optimum trouvé à 15% de couplage (en scan 5Hz, 0-10V)
    Position des lambdas :
    - L/2 en injection : 333°
    - L/4 en injection : 48°
    - L/2 en réflection : 225°
    - L/4 en réflection : 54°

- Déplacement de la photodiode en transmission. 
    * Réglage de l'iris au niveau de la photodiode : Iris ouvert PV résonance 31 mV, Iris fermée 25 mV

- Lock de la cavité.
    * transmission : 2.65V (60dB)
    * reflection : 16.5% de couplage
    * mesure devant la caméra en transmission (Power meter PH100-Si-HA-OD1 "226316" + OD2) : 0.635 uW
    * image (P.J) sauvegardé bin

Attachment 1: Screenshot_2021-12-14_2_125459.png
Screenshot_2021-12-14_2_125459.png
  48   Thu Dec 16 10:40:55 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 14 12 2021 injection optimization (and calibration)

Après-midi :
    * Optimisation de l'alignement en injection pour avoir le plus de puissance en transmission
    * Optimisation (vérification) de la polarisation pour avoir le plus de couplage.
    * tension photoDiode en réflexion (50 Ohm) : max à 136 mV (couplage 20%)
    * tension Photodiode transmission PDA100A-EC(50 ohm) : max à 3.6 V (60 dB de gain)

- Passage à 50 W en sortie de TANGOR.
    * Vérificaton de l'alignement pour avoir le plus de puissance en transmission.
    * Vérification de la polarisation pour avoir le plus de couplage.
    * tension Photodiode transmission PDA100A-EC(50 ohm) : max à 4.9 V (40 dB de gain)
    * Changement de la polarisation sur la ligne de réflexion (uniquement la lambda/4) pour ne pas saturer la photodiode.
    * tension photoDiode en réflexion (50 Ohm) : max à 136 mV (couplage 12%)
    * Références caméras :
        > NF_refl : x0 = 11.9 mm; y0 = 8.65 mm; sigma_x = 0.185 mm; sigma_y = 0.167 mm.
        > FF_refl : x0 = 2.44 mm; y0 = 0.61 mm; sigma_x = 0.241 mm; sigma_y = 0.259 mm.
        > NF_inj : x0 = 8.57 mm; y0 = 5.40 mm; sigma_x = 0.123 mm; sigma_y = 0.127 mm.

- Plus de splitter 90/10 en sortie de TANGOR, 4.5 W en sortie TANGOR :
    * Compensation du walk-of introduit par ce splitter.
    * réalignement de la cavité après locking
    * Photodiode Transmission (60dB + densité OD30) : max 2V
    * Reflection (près changement de pile) : 18% de couplage (tension max 136 mV, 50 Ohm)

 

Kevin Dupraz wrote:

Matin :

moteurs :
M1, M2 (plans): 5.110500m
M3, M4 (sphériques): -2.900000m


- Optimization de la première L/2 (en sortie tangor), le flux est maximisé en sortie du combiner.
    * mesure sortie Tangor (Power meter de puissance) : 85 + 90 mW
    * mesure sortie Tangor (Power meter PH100-Si-HA-OD1 "226316" + OD2) : 4.11mW
    * mesure avant combiner, entre miroirs de replis Newfocus (Power meter + OD2) : 3.53mW
    * mesure après pick-off (injection cavité) (Power meter + OD2) : 2.93mW

- Optimization du couplage avec la polarisation :
    * modification de la polarisation injecté dans la cavité
    * optimisation du flux reçu sur la photodiode en réflection
    * mesure du couplage.
    ** Optimum trouvé à 15% de couplage (en scan 5Hz, 0-10V)
    Position des lambdas :
    - L/2 en injection : 333°
    - L/4 en injection : 48°
    - L/2 en réflection : 225°
    - L/4 en réflection : 54°

- Déplacement de la photodiode en transmission. 
    * Réglage de l'iris au niveau de la photodiode : Iris ouvert PV résonance 31 mV, Iris fermée 25 mV

- Lock de la cavité.
    * transmission : 2.65V (60dB)
    * reflection : 16.5% de couplage
    * mesure devant la caméra en transmission (Power meter PH100-Si-HA-OD1 "226316" + OD2) : 0.635 uW
    * image (P.J) sauvegardé bin

 

Attachment 1: Screenshot_2021-12-14_3_182645.png
Screenshot_2021-12-14_3_182645.png
  49   Thu Dec 16 12:12:10 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 16 12 2021 montée en puissance

Matin: 
* Redémarrage Tangor 4W
    - couplage 15%
    - max photodiode Transmission (60dB, OD30) : 1.9V

* Montée en Puissance 20 W
    - début des effets thermiques : dérive thermique (la cavité chauffe et le piezo doit suivre, la tension baisse)
    - max photodiode Transmission (40dB, OD30) : 1.6 V
    - Reflection (~12%) ~140mV max: 
        * quand résonance => plus de signal sur la reflexion que hors lock.
        - Hypothèses: 
            saturation de la photodiode ==> avec une densité OD10 (même effet)
            effet de polarisation: essaie avec la densité OD10 (absorbante) dans le faisceau au niveau de la lambda/4 en reflection ==> la densité est brulée (ne tient pas la puissance)
            ==> essayer avec densité réflective.
        * impossible de locker sur le max de transmission

Attachment 1: Screenshot_2021-12-16_0_115846.png
Screenshot_2021-12-16_0_115846.png
  50   Thu Dec 16 12:13:52 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 16 12 2021 montée en puissance

 

Kevin Dupraz wrote:

Matin: 
* Redémarrage Tangor 4W
    - couplage 15%
    - max photodiode Transmission (60dB, OD30) : 1.9V

* Montée en Puissance 20 W
    - début des effets thermiques : dérive thermique (la cavité chauffe et le piezo doit suivre, la tension baisse)
    - max photodiode Transmission (40dB, OD30) : 1.6 V
    - Reflection (~12%) ~140mV max: 
        * quand résonance => plus de signal sur la reflexion que hors lock.
        - Hypothèses: 
            saturation de la photodiode ==> avec une densité OD10 (même effet)
            effet de polarisation: essaie avec la densité OD10 (absorbante) dans le faisceau au niveau de la lambda/4 en reflection ==> la densité est brulée (ne tient pas la puissance)
            ==> essayer avec densité réflective.
        * impossible de locker sur le max de transmission

 

Attachment 1: Screenshot_2021-12-16_1_121553.png
Screenshot_2021-12-16_1_121553.png
  51   Thu Dec 16 18:16:12 2021 BlancUnder Processinfolasers and opticsOptical roomChangements et ajout d'optiques

Optiques ajoutées :

- splitter 90/10 au début de la voie de réflexion, pour éjecter le flux et éviter que les lambdas ne chauffent.

- deuxième polariser pour avoir une meilleur extinction.

- splitter 10/90 après la première lentille sur la voie d'injection qui envoie 10% du flux sur la photodiode de réflexion.

Optiques changées :

- miroir HR à la place d'un splitter 97/3 à 0° que l'on utilisait à 45° dont le faisceau en transmission allait sur la photodiode de réflexion.

- splitter 70/30 à la place  d'un splitter 98/2 à 0° que l'on utilisait à 45° qui divisait champ proche/champ lointain.

Mise en place d'un dump pour le faisceau réfléchi par le splitter 90/10 du début de la voie de transmission.

  52   Fri Dec 17 17:03:34 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 17 12 2021 reprise réf nouvelle injection pour polarisation

Tangor : 4.2W
* Réalignement de l'injection suite aux modifications :
    - Changement de l'ordre des waveplate maintenant : lambda/4, puis lambda/2
    - suppression de l'iris au niveau de la transmission, perte d'au moins un facteur sur le signal en transmission
    - reprise des références sur les caméras en reflection. (NF:[10.2548, 6.1092], FF:[1.9881, 1.5929])

* Optimisation du couplage en pointé, puis en polarisation
    --> Maximum de reflection dans la photodiode (transmission maximale)
    - couplage mesuré en réflection : 32% +- 4%(PV) max (180 mV)
    - max intensité en transmission (0 dB, sans densité) : 1.1V

Tangor : 20W
    - Montée en puissance: grosse dérive thermique mais le signal en réflection est "correcte".
    - transmission (30dB, OD30) : pic max à 2.6 V
                    "puissance stable", 1.88V
    - couplage : 25%  (40mV/160mV)

On part en vacances ^^

Attachment 1: Screenshot_2021-12-17_0_164132.png
Screenshot_2021-12-17_0_164132.png
Attachment 2: Screenshot_2021-12-17_1_170202.png
Screenshot_2021-12-17_1_170202.png
  55   Thu May 19 10:04:05 2022 BlancUnder Processreportlasers and opticsOptical roomChangement cadence et duty cycle + ref cam

- Prise des nouvelles ref caméras : NF{709;371} , FF{389;280} [Image 1 & 2]

- Placement d'une photodiode rapide en transmission pour voir le train d'impulsions du Tangor à travers la cavité.

- Lock du Tangor en quasi-continue, P_moy = 2W, 100 kHz.

- Baisse de la cadence à 10 kHz, duty cycle 50% (sur le soft), 25 % (sur l'oscillo), à comprendre, P_moy = 1W [Image 3,4,5]

- On observe que la forme du burst a un impact sur le PDH et par consequent sur le piezo [Image3 & 4]

- Possibilité d'ajuster cela en jouant sur le gain du PID, mais si on baisse en cadence, en duty cycle et qu'on monte en puissance cela suffira ?

Attachment 1: Ref_NF_inj_18_05_2022.png
Ref_NF_inj_18_05_2022.png
Attachment 2: Ref_FF_inj_18_05_2022.png
Ref_FF_inj_18_05_2022.png
Attachment 3: Screenshot_2022-05-18_2_165654.png
Screenshot_2022-05-18_2_165654.png
Attachment 4: Screenshot_2022-05-18_5_173440.png
Screenshot_2022-05-18_5_173440.png
Attachment 5: Screenshot_2022-05-18_6_174947.png
Screenshot_2022-05-18_6_174947.png
  56   Tue Jun 14 11:16:37 2022 BlancUnder Processissuelasers and opticsOptical roomProblèmes Cavitée

Précédentes ref pour lesquelles la manip fonctionnait :

Caméras :

709 ; 371 NF

389 ; 280 FF

Moteurs :

5.091422 ; 5.091422 ; -2.9 ; -2.9

Problèmes dans la cavité :

- Pas de beating de mode alors que l'on est sur les références d'injection

- Faisceau carré en transmission

- Les softs étaient ouverts et la clé de la lockline tournée en position laser alors que j'avais tout fermé à la fin des manips et mis en standby...

 

Attachment 1: Transmission.PNG
Transmission.PNG
Attachment 2: NF_inj.PNG
NF_inj.PNG
Attachment 3: FF_inj.PNG
FF_inj.PNG
  57   Tue Jan 10 18:04:20 2023 BlancFixedinfolasers and opticsOptical roomDonnées pour la baisse en cadence

Position Cam :

692;363 NF_inj
520;193 FF_inj

Position Moteurs :

5.085435
5.085435
-2.9
-2.9

Spectre de l'oscillateur ~6 nm largeur (correct, cf image)

Durée d'impulsion : 3.35 ps (cf image)

On peut également voir que la forme du mode TEM00 a changé (cf image sur la caméra de transmission), dû à un endommagement d'un/des miroirs de la cavité. On voit également que rien n'est visible sur la caméra de réflection.

AOM 100 Hz , t_burst = 5 µs, PP 50 kHz : P_burst = 0.15 W, P_burst_PhD = 0.47 mW

AOM 100 Hz , t_burst = 5 µs, PP 50 kHz : P_burst = 0.25 W, P_burst_PhD = 0.75 mW

AOM 100 Hz , t_burst = 5 µs, PP 50 kHz : P_burst = 0.8 W, P_burst_PhD = 2.09 mW

J'ai également mis les mesures de long-term, ainsi que les burst correspondants.

Attachment 1: Spectre_oscillateur_GHz_10_01_2023.PNG
Spectre_oscillateur_GHz_10_01_2023.PNG
Attachment 2: Pulse_Width_Compressor_10000µm_10_01_2023.png
Pulse_Width_Compressor_10000µm_10_01_2023.png
Attachment 3: Mode_TEM00_lockline_lockée_après_cassage_cavité_trans2_10_01_2023.PNG
Mode_TEM00_lockline_lockée_après_cassage_cavité_trans2_10_01_2023.PNG
Attachment 4: Mode_TEM00_lockline_lockée_après_cassage_cavité_trans_10_01_2023.PNG
Mode_TEM00_lockline_lockée_après_cassage_cavité_trans_10_01_2023.PNG
Attachment 5: Mode_TEM00_lockline_lockée_après_cassage_cavité_NF_refl_10_01_2023.PNG
Mode_TEM00_lockline_lockée_après_cassage_cavité_NF_refl_10_01_2023.PNG
Attachment 6: Long-term_Histo.png
Long-term_Histo.png
Attachment 7: Tangor_and_Cavity_burst.png
Tangor_and_Cavity_burst.png
  58   Wed Jan 18 17:52:33 2023 BlancFixedinfolasers and opticsOptical roomTests et mesures lockline

Mesures de puissance fibres lockline :

AVANT POLISSAGE

injection premier étage : 15 mW

sortie premier étage : 63 mW

Sortie AOM à 250 MHz : 2.5 mW

Sortie EOM : 1.3 mW

Sortie : 20 =>40 mW

APRES POLISSAGE

injection premier étage : 15 mW

sortie premier étage : 63 mW

Sortie AOM à 159 MHz : 2.9 mW            Sortie AOM à 250 MHz : 15 mW

Sortie EOM : 1.5 mW                              Sortie EOM : ?

Sortie : 47 mW                                        Sortie : 54 mW

 

Même après polissage de la fibre de sortie du premier étage, on observe que la puissance diminue d'un facteur 4 à travers l'AOM drivé à sa fréquence nominale.

  59   Thu Jan 19 14:56:27 2023 BlancFixedinfolasers and opticsOptical roomTests et mesures lockline

 

Blanc wrote:

Mesures de puissance fibres lockline :

AVANT POLISSAGE

injection premier étage : 15 mW

sortie premier étage : 63 mW

Sortie AOM à 250 MHz : 2.5 mW

Sortie EOM : 1.3 mW

Sortie : 20 =>40 mW

APRES POLISSAGE

injection premier étage : 15 mW

sortie premier étage : 63 mW

Sortie AOM à 159 MHz : 2.9 mW            Sortie AOM à 250 MHz : 15 mW

Sortie EOM : 1.5 mW                              Sortie EOM : ?

Sortie : 47 mW                                        Sortie : 54 mW

 

Même après polissage de la fibre de sortie du premier étage, on observe que la puissance diminue d'un facteur 4 à travers l'AOM drivé à sa fréquence nominale.

 

Attachment 1: AOM_250_drivé_à_160_MHz_sortie_fibre_bleue.png
AOM_250_drivé_à_160_MHz_sortie_fibre_bleue.png
Attachment 2: schéma_mesures_lockline_18_01_2023.PNG
schéma_mesures_lockline_18_01_2023.PNG
  60   Thu Jan 19 15:05:58 2023 BlancFixedinfolasers and opticsOptical roomSuite Mesures de puissance lockline

Changement d'AOM pour passer sur l'AOM 200 MHz.

Le schéma suivant récapitule les mesures prises.

Il y a eu une chute de la puissance en sortie de la fibre bleue de la lockline.

1) à la vue des mesures, il semble que le deuxieme etage de la lockline soit bien fatigué

2) La fibre bleue s'est déteriorée d'un seul coup...

Par contre il n'y a plus les modulations observées précédement sur la photodiode grâce au changement d'AOM, à la fois à 159 MHz et 200 MHz.

 

Attachment 1: schéma_mesures_lockline_19_01_2023.PNG
schéma_mesures_lockline_19_01_2023.PNG
Attachment 2: 20230119_124318.jpg
20230119_124318.jpg
  61   Thu Jan 19 18:20:02 2023 BlancFixedinfolasers and opticsOptical roomSuite Mesures de puissance lockline

 

Blanc wrote:

Changement d'AOM pour passer sur l'AOM 200 MHz.

Le schéma suivant récapitule les mesures prises.

Il y a eu une chute de la puissance en sortie de la fibre bleue de la lockline.

1) à la vue des mesures, il semble que le deuxieme etage de la lockline soit bien fatigué

2) La fibre bleue s'est déteriorée d'un seul coup...

Par contre il n'y a plus les modulations observées précédement sur la photodiode grâce au changement d'AOM, à la fois à 159 MHz et 200 MHz.

 

Repolissage de la fibre bleue et de la fibre de sortie du deuxieme étage.

Fin des manips :

sortie deuxieme étage : 60 mW

sortie fibre bleue : entre 50 et 55 mW

  62   Fri Jan 20 18:01:22 2023 BlancFixedinfolasers and opticsOptical roomMesure de finesse

Fréquence centrale : 879.48 MHz

FWHM mesurée à la main sans sweep : 120 / 130 KHz

=> Finesse = FSR/FWHM = ~7000

Apporter un analyseur de spectre pour régler le sweep du Marconi pour une mesure de finesse plus précise.

  63   Wed Jan 25 12:12:36 2023 BlancFixedinfolasers and opticsOptical roomMesure de finesse

 

Blanc wrote:

Fréquence centrale : 879.48 MHz

FWHM mesurée à la main sans sweep : 120 / 130 KHz

=> Finesse = FSR/FWHM = ~7000

Apporter un analyseur de spectre pour régler le sweep du Marconi pour une mesure de finesse plus précise.

On a pu mesurer la FWHM à l'aide du sweep : 124kHz

On obtient une finesse de 7300 en ayant traité les données.

Attachment 1: Mesure_de_finesse.png
Mesure_de_finesse.png
  64   Wed Jan 25 12:18:25 2023 BlancFixedinfolasers and opticsOptical roomPosition des miroirs et gain photodiode

Nouvelle position des miroirs :

5.198865 ; 5.199305 ; -2.9 ; -2.9

Photodiode de transmission, gain de 60, 1MOhm osc : 2.5 V

Photodiode de transmission, gain de 50, 1MOhm osc : 800 mV

  65   Thu Jan 26 22:49:46 2023 BlancFixedinfolasers and opticsOptical roomNouvelles positions moteurs et calibrations camera

Positions moteurs :

5.198865 ; 5.198596 ; -2.9 ; -2.9

Calibration caméras :

NF_inj : 957;616

FF_inj : 328;2790

On a réussi à avoir un signal d'erreur très proche de la théorie, ainsi qu'un lock très propre. (Cf images)

En regardant laTF de la lockline, on a vu la modulation à +ou- 159 MHz autour du pic à 880 MHz de l'oscillateur GHz, mais cela n'explique pas l'offset sur le PDH.

Hypohèse à vérifier : le câble d'alimentation du PDH qui est défectueux.

Attachment 1: Lock_lockline.png
Lock_lockline.png
Attachment 2: Signal_d'erreur_PDH.png
Signal_d'erreur_PDH.png
  66   Mon Jan 30 14:39:52 2023 BlancFixedinfolasers and opticsOptical roomMesures de Polarisation

En récupérant la transmission du miroir M3 sphérique sortant de la cavité, on a pu faire une mesure de polarisation de la cavité. (Cf courbes ci-dessous)

La moyenne pour chaque courbe donne :

S0 = 1.0058 ; S1 = 0.4643 ; S2 = 0.0488 ; S3 = 0.8913

DOP = 1.0062 ; DOLP = 0.4669 ; DOCP = 0.8913

Phase = 86.8705° ; Ellipticité = 0.6076 ; Êta = 2.9962

Température de la tête du polarimètre : 39.9815

Par reconstruction, on obtient l'ellipse donnée en bleu ci-dessous.

Attachment 1: Paramètres_de_Stokes.png
Paramètres_de_Stokes.png
Attachment 2: DOP_DOLP_DOCP.png
DOP_DOLP_DOCP.png
Attachment 3: Phase_Ellipticité_Êta.png
Phase_Ellipticité_Êta.png
Attachment 4: Ellipse_polarisation_cavité.png
Ellipse_polarisation_cavité.png