  Status of commissioning, report also here plots are reports., Page 3 of 5  Not logged in ELOG logo
    Reply   Mode observed and cavity Lock , posted by Manar Amer at Other about lasers and optics 20212804_Fundamental_mode_in_cavity_2D_-_Position.png20212804_Cavity_Coupling_with_Koheras.jpg20212804_Temperature_Curve.jpg
** Following up from yesterday an observation about the TM00 mode : it has been seen with similar dimensions and its position is the same, this was before
restarting the temperature regulator. 

                 --> after restarting the temp. regulator 
    Reply   Mode observed and cavity Lock , posted by Manar Amer at Other about lasers and optics NF_Trans__2021-06-25__16-38-56.bmp
New configuration is made. The new mode observed in NF_Trans is the following :




Manar
Amer wrote:



** Following up from yesterday an observation about the TM00 mode :
Entry   Mirrors Mottors Connections , posted by Manar Amer at Other about mechanics 20210804_Motor_connections_with_mirrors_inside_cavity_box.jpg20210804_Mottor_cables_behind_cavity_box.jpg
* The connections to control the motors of the mirrors are connected in the order stated :

                 M1       
-      M2       -       M3     -   
    Reply   Mirrors Mottors Connections , posted by Manar Amer at Other about mechanics | cabling 
Added the cabling to the categories 




Manar
Amer wrote:



* The connections to control the motors of the mirrors are connected
Entry   Mesures de Polarisation Lockline et Tangor, posted by Blanc at Optical room about lasers and optics 11x
Vérification des polarisations après le combiner pour le Tangor et la Lockline.

On remarque certains paramètres de Stokes ont des valeurs étranges, comme S1 de la lockline.

De plus le polariseur voit la lockline comme non-polarisé.
Entry   Mesures de Polarisation, posted by Blanc at Optical room about lasers and optics Paramètres_de_Stokes.pngDOP_DOLP_DOCP.pngPhase_Ellipticité_Êta.pngEllipse_polarisation_cavité.png
En récupérant la transmission du miroir M3 sphérique sortant de la cavité, on a pu faire une mesure de polarisation de la
cavité. (Cf courbes ci-dessous)

La moyenne pour chaque courbe donne :
    Reply   Mesures de Polarisation, posted by Blanc at Optical room about lasers and optics Ellipse_Polarization_CAVITY.png
Correction su l'orientation de l'ellipse.




Blanc
wrote:



En récupérant la transmission du miroir M3 sphérique
Entry   Mesure polarisation voie de réflexion, posted by Blanc at Optical room about lasers and optics 18x
J'ai fait les mesures de polarisation sur la voie de réflexion pour voir la polarisation avant le glan et après le glan du Tangor et
de la lockline.

Dans l'ordre :
Entry   Mesure de finesse, posted by Blanc at Optical room about lasers and optics 
Fréquence centrale : 879.48 MHz

FWHM mesurée à la main sans sweep : 120 / 130 KHz

=> Finesse = FSR/FWHM = ~7000
    Reply   Mesure de finesse, posted by Blanc at Optical room about lasers and optics Mesure_de_finesse.png
 




Blanc
wrote:



Fréquence centrale : 879.48 MHz
Entry   Manip 17 12 2021 reprise réf nouvelle injection pour polarisation, posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-17_0_164132.pngScreenshot_2021-12-17_1_170202.png
Tangor : 4.2W
* Réalignement de l'injection suite aux modifications :
    - Changement de l'ordre des waveplate maintenant : lambda/4, puis lambda/2
Entry   Manip 16 12 2021 montée en puissance, posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-16_0_115846.png
Matin: 
* Redémarrage Tangor 4W
    - couplage 15%
    Reply   Manip 16 12 2021 montée en puissance, posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-16_1_121553.png
 




Kevin
Dupraz wrote:



Matin: 
Entry   Manip 14 12 2021 injection optimization (and calibration), posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-14_2_125459.png
Matin :

moteurs :
M1, M2 (plans): 5.110500m
    Reply   Manip 14 12 2021 injection optimization (and calibration), posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-14_3_182645.png
Après-midi :
    * Optimisation de l'alignement en injection pour avoir le plus de puissance en transmission
    * Optimisation (vérification) de la polarisation pour avoir le plus de couplage.
Entry   Manip 10 12 2021 (optimization injection Burst), posted by Kevin Dupraz at Optical room about lasers and optics NF_Refl__2021-12-10__14-25-08.pngtest__2021-12-10__14-25-08_1920_25gm.png
Manip 10/12/2021
Après recherche de la résonnance principale
    - MCS miroirs plan à 5.110250mm (en derive vers les negatifs)
Entry   Locking with fast analog loop on AOM, posted by Ronic Chiche at Optical room about lasers and optics 
in addition to the standard locking scheme with the GHz laser PZT,

we added an AOM after the PDH modulation EOM and we drove it with an FM modulated signal generator (FMDev = 2.4MHz) seeded by the error signal.
(we didn't put a 50ohm plug to adapt the error signal coming from the PDH box, otherwise, it is too much smaller)
Entry   Locking of the cavity and setup changing, posted by Blanc at Optical room about lasers and optics | detectors and electronics scope_0.png
We change the setup a bit to have the same polarization on the photodiode and the PDH (in reflexion of the cavity). CF: schematics (futur).

Thanks to that, the symetry between reflexion and transmission is better now (as you can see on the plot).

We also put more power on PHD to have a better SNR by changing splitters. (We had 100 µW and now we have more than 1 mW, at the expense
Entry   Lock, cavity length and finesse estimate, posted by Aurélien Martens at Optical room about lasers and optics | detectors and electronics IMG_20210430_173331.jpg
Ronic just suceeded in obtaining a reasonably good lock on the cavity. Air flow is switched off.

We stopped all movements (closed loop, click STOP in the PTC interface) and saw immediately a fair but not excellent lock.

We then switched off the smaract motors and the obtained lock was good. Switching on again the smaract means tthat the references are lost.
    Reply   Lock, cavity length and finesse estimate, posted by Aurélien Martens at Optical room about lasers and optics | detectors and electronics 
we observed that the transmitted light coming from the cavity is made of the cavity mode light and the scattered light on the borders of the mirrors.
the scattered light on the borders of the mirrors is triggered by the cavity mode itself when the cavity is locked.
we didn't put an iris to cut this scattered light coming from the cavity.