HOME SBOX THOMX MINICAV Utilities
THOMX orders THOMX installation THOMX commissioning THOMX control command
  Status of commissioning, report also here plots are reports., Page 22 of 23  Not logged in ELOG logo
ID Date Author Status Type Category Location Title
  32   Mon Apr 8 11:55:40 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomP4 cleaned and finesse measurement

P4 cleaned with isopropanol.

Finesse measured afterward: ~=13 000 (clear increase)

  31   Mon Apr 8 10:18:31 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomFinesse after 2days

Finesse measured this morning after P1 SUPRASIL being into the cavity for two days. In order to check if it's getting dirty.

Finesse still ~= 11 000

  30   Fri Apr 5 11:28:39 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomFinesse with P1 SUPRASIL

P1 SUPRASIL has been installed instead of P1 Sapphire. It has been cleaned 3 times on each face with wipe+isopropanol before to be installed.

Finesse ~= 10 000 - 11 000 (EOM Sweep of 100 kHz in 10sec, see last image.)
Coupling ~= 45%

Motors position:
P1 = -825 000
S2 = +825 000
S3 = +825 000
P4 = -825 000

  29   Fri Apr 5 11:18:47 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomTransmission of the mirrors

P1 transmission out of the mount (means angle of the mirror is not the real angle):

Pin = 3.2 mW
Pout = 545 nW
Offset = 68 nW

Transmission = (545-68)/3.2e6 = 149 ppm

Loïc Amoudry wrote:

After verification, transmission of P1 correspond to ~150ppm.

The 1337ppm come from a beam leakage. Reflection from P1 goes to the input window which reflects again a part of the beam directly into the cavity between P1 and S3 then between P4 and S2.

Loïc Amoudry wrote:

We openned the cavity, removed S2 and measured trans of P1 and same for S2.

Pin ~= 3.91 mW
PS2 ~= 20 nW --> ~~ 5 ppm !!!
PP1 ~= 5.23 uW --> ~~ 1337 ppm !?!?!

P1 is the Sapphire mirror. 3mm thick. Tomorrow we'll measure its trans again, being out of the mount for constraint issues.

 

 

  28   Thu Apr 4 09:42:12 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomTransmission of the mirrors

After verification, transmission of P1 correspond to ~150ppm.

The 1337ppm come from a beam leakage. Reflection from P1 goes to the input window which reflects again a part of the beam directly into the cavity between P1 and S3 then between P4 and S2.

Loïc Amoudry wrote:

We openned the cavity, removed S2 and measured trans of P1 and same for S2.

Pin ~= 3.91 mW
PS2 ~= 20 nW --> ~~ 5 ppm !!!
PP1 ~= 5.23 uW --> ~~ 1337 ppm !?!?!

P1 is the Sapphire mirror. 3mm thick. Tomorrow we'll measure its trans again, being out of the mount for constraint issues.

 

  27   Tue Apr 2 18:05:46 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomBeam size at closest spherical mirrors position

S3 and S2 @ +1 640 000 and P1 and P4 @ -825 000

Diameter in x: 4.7mm
Diameter in y: 4.7mm

  26   Tue Apr 2 18:03:31 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomTransmission of the mirrors

We openned the cavity, removed S2 and measured trans of P1 and same for S2.

Pin ~= 3.91 mW
PS2 ~= 20 nW --> ~~ 5 ppm !!!
PP1 ~= 5.23 uW --> ~~ 1337 ppm !?!?!

P1 is the Sapphire mirror. 3mm thick. Tomorrow we'll measure its trans again, being out of the mount for constraint issues.

  25   Tue Apr 2 17:59:18 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomFinesse after removing wires

19/04/01: We removed the wires from the beam path in the cavity. Finesse stayed at 3400. FSR: 33.342 MHz.

When locked:

Input Power: ~4mW
Through S2: ~0.950uW
Through S3: ~0.930uW
Through P4: ~0.850uW

  24   Fri Mar 29 18:33:15 2019 Ronic ChicheFixedissuemechanics | lasers and opticsOptical roomAlignment procedure

we did again the complete alignment procedure starting with iris and optimizing injection motors Ma and Mb, then installing mirror S2, S3, P4 and P1, optimizing thetaX and thetaY axis for each motor.

everything was fine until we installed P1. we tried to optimize thetaX and thetaY of P1 and we clearly observed a strange motion when doing that :
   - for thetaX axis, steps in one direction seem to have a different length from steps in reverse direction.
   - for thetaY axis, trying to move in one direction, makes sometime a motion in the reverse direction.
at this moment, it is difficult to say if the problem comes from the controller, the mirror mount or the motor itself.
if both axis are concerned on the same mirror (P1), maybe the problem comes from the mount... to be continued on Monday.

 

  23   Thu Mar 28 09:25:58 2019 Loïc AmoudryFixedissuemechanicsOptical roomMotors Z axis reversed

All Z motors (for P1, S2, S3 and P4) have been initialized again with a proper HOME command with their own controller  with the "superviser" software (which means they all moved to the +1650000 position).
Thus, they have been moved to their default position : +825000 for spherical mirrors (S2 and S3) and -825000 for plan mirrors (P1 and P4).

reminder : lower values on the motors makes the motor tip going inside which makes the cavity length larger.
thus spherical mirrors are in default position with a possible big range to put inside the motor tip (for thermal effects compensation).
by complementarity, plan mirrors are in default position with a possible big range to put outside the motor tip.

the Labwindows CVI was buggy about the proper representation of the motors motion in Z direction (wrong direction and wrong default position).
the software is now fixed.
 

Loïc Amoudry wrote:

S2 and S3 z-axis are initially at 1/4 (5mm/20mm) of there range over ~20mm range. This is meant to be able to spread them up to ~15mm in case of thermal effect.

Actually, they are in an opposite position. Means that they can only be spread of 5mm from the initial position.

 

  22   Tue Mar 26 18:32:40 2019 Loïc AmoudryFixedreportmechanics | lasers and optics | cablingOptical roomWires in laser path

There was some ceramic wires close to the beem path right before P4. We moved them away (we can imagine they could move again with vibration).

  21   Tue Mar 26 18:26:32 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomCheck of mirrors and coating directions

Direction of HR face of P1 and P4 have been checked, there was no problem. The mirrors are in the good direction and the coating is on the good face.

Once P1 and P4 outside of the cavity, we checked the coating position by sending the HeNe laser with an angle. There was two reflections and the most intense was the one closer to the normal to the mirror.

On P4 we noticed a spot which seems to be in the coating (does not move while sending nitrogen).

  20   Fri Mar 22 17:53:10 2019 Loïc AmoudryFixedreportlasers and optics | vacuumOptical roomCavity at atm pressure

Dry air injected into the cavity.

Lock has been proceeded without any trouble. ~~4000 finesse. Considered to be the same than before.

x and y motors axis put back to the last "cavity at air" position. z position kept at maximum distance between spherical mirrors.

  19   Tue Mar 19 17:01:52 2019 Loïc AmoudryFixedissuemechanicsOptical roomMotors Z axis reversed

S2 and S3 z-axis are initially at 1/4 (5mm/20mm) of there range over ~20mm range. This is meant to be able to spread them up to ~15mm in case of thermal effect.

Actually, they are in an opposite position. Means that they can only be spread of 5mm from the initial position.

  18   Tue Mar 19 16:35:32 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomBeam size and finesse

FSR initiale : 33.34 MHz (S3 = -825000 et S4 = -825000)

Beam size initiale : wx=2.05mm et wy=2.15mm

Finesse initiale : 3400

 

FSR après éloignement des miroirs sphériques : 33.29 MHz (S3 ~= -1496000 et S4 ~= -1496000)

Beam size : wx=1.7mm et wy=1.85mm

Finesse : 3600

 

FSR après rapprochement des miroirs sphériques : 33.39 MHz (S3 ~= -250000 et S4 ~= -500000)

Beam size : wx = 2.1mm et wy=2.4mm

 

Fit de la caméra jamais au dessus de 90%. ~~80%.

  17   Tue Feb 19 15:50:08 2019 Loïc AmoudryFixedinfomechanicsOptical roomMotors connexions

Motors are connected. The true mirror's name are P1 for injection mirrors, S2, S3 and P4. (S=spherical, P=planar).

They are still some old schemes (and notations on the cavity) which mention M1, M2, M3 and M4. I make the link here:

M3 = P1
M1 = S2
M2 = S3
M4 = P4

The wires are plugged as follows (I use notations wrote on the cavity and wires):

M3x = 1-1
M3y = 1-2
M3z = 2-1

M1x = 1-3
M1y = 1-4
M1z = 2-2

M2x = 1-5
M2y = 1-6
M2z = 2-3

M4x = 1-7
M4y = 1-8
M4z = 2-4

  16   Thu Feb 14 18:20:52 2019 Loïc AmoudryFixedreportmechanics | lasers and opticsOptical roomBasic alignment & alignment motors check

14/02/19: The 4 motors but Mby control works. Mby motor home does not bring it to its "middle" position but to the end of its range. Issue circumvented using the higher rights software.

The beam is now going through the cavity windows (from P1 to S2).

  15   Fri Feb 1 14:10:25 2019 Loïc AmoudryFixedreportlasers and opticsOtherMirrors sent to LMA

30/01/19 - As LMA people came, we discussed about the spots on ThomX mirrors which should be clean. As mentionned in "SBOX commissioning", these spots are normal.

An other question appeared during the conversation about the storage time and location of these mirrors. They have been stored for more than two years in plastic boxes. LMA people explained that some company do not let anything more than 6 months in these boxes for pollution issues.

We merged into the same conclusion, clean the mirrors in case of pollution. Finally, they took the mirrors and we will go to LMA to recover them between 11/02 and 14/02. 

  14   Fri Feb 1 13:39:21 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomMirrors checked with binocular microscope

29/01/19 - ThomX mirrors have also been checked with binocular microscope. They shown the same spots as S-BOX mirrors. 

1st image shows reflective face of a spherical mirror.

2nd image shows reflective face of M1 (sapphire).

  13   Fri Feb 1 13:20:47 2019 Loïc AmoudryFixedreportutilitiesOptical roomEnd of cleaning, plastic fixed on cavity

08/02/19 - End of the cleaning process.

ELOG V3.1.4-395e101