Entry   33MHz (FP-cavity vs RF) phase stability, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics 
    Reply   33MHz (FP-cavity vs RF) phase stability, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics 
Message ID: 314     Entry time: Tue Jan 16 12:26:12 2024     Reply to this: 320
 Author: Ronic Chiche 
 Status: Fixed 
 Type: info 
 Category: lasers and optics | detectors and electronics 
 Location: ThomX igloo 
 Title: 33MHz (FP-cavity vs RF) phase stability 

this morning, with Vincent Chaumat, we made a beating (phase detection) between the laser 33MHz and the 33MHz coming from the Rigol generator locked by the 10MHz link onto the 500MHz ring generator.

once the laser and the FP-cavity are locked on the RF reference, the 33MHz phase detection is stabilized and confirm the laser/FP-cavity lock on the RF reference.

we plan to make a drift measurement and a jitter measurement to estimate this lock quality.
the 33MHz frequency being quite small, the quality of this estimation will be poor but it can give a "worse case" estimation.