Message ID: 239     Entry time: Thu Sep 7 18:51:43 2023
 Author: Ronic Chiche 
 Status: Fixed 
 Type: issue 
 Category: lasers and optics 
 Location: ThomX igloo 
 Title: Onefive laser modelock and Smaract CEP controller 

today, during an attempt in improving the CEP with the Smaract controller CH2, we lost the laser modelock.

at this point, we decided to reference the Smaract CH2 for CEP :
now, 0 mm is roughly the middle of the motor range.
the endpoints of the motor are : +/- 5.9 mm
once we found back the laser modelock, we searched for the region where the modelock was still effective : +/- 1.8 mm