Message ID: 128     Entry time: Fri Mar 11 12:46:58 2022
 Author: Ronic Chiche 
 Status: Fixed 
 Type: info 
 Category: lasers and optics 
 Location: ThomX igloo 
 Title: FP cavity mode axis change and Finesse measurement 

this morning with Manar, we did a cavity mode axis change.

we checked the cavity mode centroid position on the beam profiler (placed behind P4) and changed it by roughly 1mm on the X-axis and by 500µm on the Y-axis.

for doing that, we played only on the S2 mirror while the cavity is locked and we slightly changed X and Y motors, step by step, and we realigned the laser beam with external injection mirrors when needed.

in the end, we were able to improve the transmission power by roughly 25% (810mV to 1020mV on the Transmission photodiode on the scope).

we did several Finesse measurements after that: 4950, 5190, 5100, 5120, 5200 => ~ 5100

compare to the previous Finesse value, around 4200, this is also roughly 25% better, in agreement with the transmission power increase.

conclusion: it seems the Finesse, thus the losses are almost the same "everywhere" on the mirrors => we need to clean them or replace them.

 