HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports., Page 9 of 13  Not logged in ELOG logo
ID Date Author Status Type Category Location Title
  85   Mon May 9 19:27:35 2022 Manar AmerFixedreportlasers and optics | utilitiesOptical roomSpincoater - Mirrror cleaning

PowerPoint shows the detailed images of all the mirrors before and after cleaning.

Images of SBox mirrors and M1 from Gamma factory attached only for the front after cleaning

the lines showing are on the back of the mirrors, they come from the spinning mount the mirrors are fixed on in the spin-coater.

Naming code :

               M#BB - M# mirror Back Before cleaning

               M#BA - M# mirror Back After cleaning

               M#FB - M# mirror Front Before cleaning

               M#FA - M# mirror Front After cleaning

               G : Gamma factory

 

Manar Amer wrote:

Images of the SBox mirrors cleaned at the same time when the cavity was aligned.

Directly from the spincoater to the SBox directly.

PowerPoint attached shows details

Manar Amer wrote:

Used the spincoater and pure water to clean dielectric mirrors which will be installed inside the SBox for directing the reflection line.

The method : place pure alchole on mirror - wipe it using white sterile tissue (in one direction) - clean with pure water in spincoater suing cotton swap .

After multible rounds, the result in the images was obtained. there is still a significant amount of dust particles (not all could be dust!!).

Not sure if the reason is the cleaning method ?? will invistegate it more !!!

Manar Amer wrote:

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

 

 

 

  84   Mon May 9 19:05:35 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Today; replaced the M1 SBox mirror with M1 from gamma factory , following the previous alignment.

Removed all mirrors and cleaned them one by one.

Started with no M1, placed M2 did its alignment through M3 pinhole, then placed M3 aligned it through M4 pinhole, then placed M4 aligned it through M1 pinhole.

Placed M1 Gamma factory at the end

Note: mirror is new was not used before, took image on microscope there was no dust in the center, but there was little on the edges (cleaned it using spin-coater)

beating of fundamental mode observed (at M2 transmission) immediately with very higher order modes, slight adjustment on M4 suppressed them.

The beam is not circular and still has some deformation, M1 was changed, so it could be a damaged point at one of the other mirrors (to be investigated tomorrow)!!

No significant damage seen on the other mirrors.

 

 

Manar Amer wrote:

Last week In FP cavity we had all SBox mirrors including the damaged M1 mirror.

The beating was observed with just a slight adjustment on M4 alignment.

In addition, it was also observed while driving the piezo of the CW laser.

In the attached image of the beam at transmission point at M2 mirror, the shape of the beam is deformed.

one of the factors could be related to the damaged spot on M1 mirror. The deformation position can also be seen in the higher order modes.

after changing the injection mirror to Gamma factory M1, this deformation should disappear.

Manar Amer wrote:

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

 

 

  83   Mon May 9 18:36:48 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Last week In FP cavity we had all SBox mirrors including the damaged M1 mirror.

The beating was observed with just a slight adjustment on M4 alignment.

In addition, it was also observed while driving the piezo of the CW laser.

In the attached image of the beam at transmission point at M2 mirror, the shape of the beam is deformed.

one of the factors could be related to the damaged spot on M1 mirror. The deformation position can also be seen in the higher order modes.

after changing the injection mirror to Gamma factory M1, this deformation should disappear.

Manar Amer wrote:

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

 

  82   Wed Apr 27 22:30:14 2022 Manar AmerFixedreportlasers and optics | utilitiesOptical roomSpincoater - Mirrror cleaning

Images of the SBox mirrors cleaned at the same time when the cavity was aligned.

Directly from the spincoater to the SBox directly.

PowerPoint attached shows details

Manar Amer wrote:

Used the spincoater and pure water to clean dielectric mirrors which will be installed inside the SBox for directing the reflection line.

The method : place pure alchole on mirror - wipe it using white sterile tissue (in one direction) - clean with pure water in spincoater suing cotton swap .

After multible rounds, the result in the images was obtained. there is still a significant amount of dust particles (not all could be dust!!).

Not sure if the reason is the cleaning method ?? will invistegate it more !!!

Manar Amer wrote:

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

 

 

  81   Wed Apr 27 22:13:37 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

  80   Tue Apr 26 20:08:10 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

  79   Mon Apr 25 18:02:09 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

  78   Mon Apr 25 17:58:01 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

  77   Mon Apr 25 17:28:29 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

  76   Wed Apr 20 19:05:37 2022 Manar AmerFixedreportlasers and opticsOptical roomKBox Alignment

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

  75   Wed Apr 13 18:42:50 2022 Manar AmerFixedreportlasers and optics | utilitiesOptical roomSpincoater - Mirrror cleaning

Used the spincoater and pure water to clean dielectric mirrors which will be installed inside the SBox for directing the reflection line.

The method : place pure alchole on mirror - wipe it using white sterile tissue (in one direction) - clean with pure water in spincoater suing cotton swap .

After multible rounds, the result in the images was obtained. there is still a significant amount of dust particles (not all could be dust!!).

Not sure if the reason is the cleaning method ?? will invistegate it more !!!

Manar Amer wrote:

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

 

  74   Fri Apr 8 19:27:12 2022 Manar AmerFixedissuedetectors and electronics | utilitiesOptical roomClean Airflow Dust measurements

Dust measurement done today on  top of the SBox average 10min

Manar Amer wrote:

The dust meter cap was cleaned using Alcohol, and using the filter white cap the dust count was (0 Av. 10 min)

After the dust counter was hand held in direct airflow, it counted (1581 p/m3 Av. 2 min).

I tested the count also in the SAS, and it counted twice the amount ~ 4500 p/m3

Note: discussion on the next steps to take for the airflow filtering !!!!!!!!

 

  73   Mon Apr 4 13:58:23 2022 Manar AmerFixedissuedetectors and electronics | utilitiesOptical roomClean Airflow Dust measurements

The dust meter cap was cleaned using Alcohol, and using the filter white cap the dust count was (0 Av. 10 min)

After the dust counter was hand held in direct airflow, it counted (1581 p/m3 Av. 2 min).

I tested the count also in the SAS, and it counted twice the amount ~ 4500 p/m3

Note: discussion on the next steps to take for the airflow filtering !!!!!!!!

  72   Thu Mar 31 13:29:01 2022 Manar AmerFixedreportmechanics | lasers and optics | utilitiesOptical roomDust Cleaning of SBox

After an overnight where the cavity box was left open, a measurement of the Dust meter was taken:

Outside on the optical table on direct air flux (1701 p/m3 Av 10 min) and

Inside the cavity box (998 p/m3 Av 10 min)

The dust inside the cavity has decreased by ~ 30 % for 0.3 u particles.

Note: the cavity box is still open.

Manar Amer wrote:

The inside of the cavity was cleaned today. It was very dirty and had big dust particles and fibers inside.

Cleaning was done for everything; wall, floor, fixed mounts, wires (still have some dust difficult to remove), edges.

The dust meter average was measured outside the cavity (2471 p/m3 av 6 min)  and inside after cleaning (1424 p/m3 av 10 min).

The cavity was left open under the clean air flow, will observe it under UV, clean and measure Dust tomorrow.

Images attached are after cleaning.

Objects inside the cavity at the moment :

  • 4 main mirror mounts
  • 4 mirror fixers on mounts and 3 screws each (total 12)
  • 2 reflecting mirror mounts
  • 2 dielectric mirrors (mounted and to be cleaned)
  • two mounts for D-shaped mirrors.
  • 2 D-shaped mirrors (unmounted, and they will need cleaning)
  • 2 metal irises (pinholes)

 

Manar Amer wrote:

The cleaning was done twice with half an hour interval between the two times
reading on the dust meter was taken after each time, and it was zero averaged over a 10-min

Manar Amer wrote:

An Initial cleaning of the KBoX outside was done, All the optics and mechanics placed that were on the table were removed.

Only the old Mitylaser amplifier remains on the table.

Since we will be using the ThomX Celia amplifier, we will remove it and place it in a secure location (to be decided).

Tomorrow a cleaning of the inside of the cavity box is to be done.

 

 

 

  71   Wed Mar 30 18:53:01 2022 Manar AmerFixedreportmechanics | lasers and optics | utilitiesOptical roomDust Cleaning of SBox

The inside of the cavity was cleaned today. It was very dirty and had big dust particles and fibers inside.

Cleaning was done for everything; wall, floor, fixed mounts, wires (still have some dust difficult to remove), edges.

The dust meter average was measured outside the cavity (2471 p/m3 av 6 min)  and inside after cleaning (1424 p/m3 av 10 min).

The cavity was left open under the clean air flow, will observe it under UV, clean and measure Dust tomorrow.

Images attached are after cleaning.

Objects inside the cavity at the moment :

  • 4 main mirror mounts
  • 4 mirror fixers on mounts and 3 screws each (total 12)
  • 2 reflecting mirror mounts
  • 2 dielectric mirrors (mounted and to be cleaned)
  • two mounts for D-shaped mirrors.
  • 2 D-shaped mirrors (unmounted, and they will need cleaning)
  • 2 metal irises (pinholes)

 

Manar Amer wrote:

The cleaning was done twice with half an hour interval between the two times
reading on the dust meter was taken after each time, and it was zero averaged over a 10-min

Manar Amer wrote:

An Initial cleaning of the KBoX outside was done, All the optics and mechanics placed that were on the table were removed.

Only the old Mitylaser amplifier remains on the table.

Since we will be using the ThomX Celia amplifier, we will remove it and place it in a secure location (to be decided).

Tomorrow a cleaning of the inside of the cavity box is to be done.

 

 

  70   Tue Mar 29 16:27:34 2022 Manar AmerFixedreportutilitiesOptical roomSpincoater - Mirrror cleaning

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

  69   Tue Mar 29 16:23:00 2022 Manar AmerFixedreportlasers and optics | cablingOptical roomMightylaser Amplifier

Migthylaser amplifier has been moved from the SBox table to the PLIC table.

 

  68   Wed Mar 23 18:17:45 2022 Manar AmerFixedreportutilitiesOptical roomDust Cleaning of SBox

The cleaning was done twice with half an hour interval between the two times
reading on the dust meter was taken after each time, and it was zero averaged over a 10-min

Manar Amer wrote:

An Initial cleaning of the KBoX outside was done, All the optics and mechanics placed that were on the table were removed.

Only the old Mitylaser amplifier remains on the table.

Since we will be using the ThomX Celia amplifier, we will remove it and place it in a secure location (to be decided).

Tomorrow a cleaning of the inside of the cavity box is to be done.

 

  67   Wed Mar 23 18:11:22 2022 Manar AmerFixedreportutilitiesOptical roomDust Cleaning of SBox

An Initial cleaning of the KBoX outside was done, All the optics and mechanics placed that were on the table were removed.

Only the old Mitylaser amplifier remains on the table.

Since we will be using the ThomX Celia amplifier, we will remove it and place it in a secure location (to be decided).

Tomorrow a cleaning of the inside of the cavity box is to be done.

  66   Mon Jul 5 14:46:44 2021 Manar AmerFixedreportlasers and optics | detectors and electronicsOptical roomOEwaves CW Laser - Phase Noise

correction on unit 1030 nm

Manar Amer wrote:

a correction on the wavelength of the laser  it is 1030 um 

Manar Amer wrote:
  • We started measuring the phase noise on the OEwaves CW laser.
    • Class 3b
    • wavelength 1.5 um
  • The procedure is done using self coupling of the laser
    • splitter 50%-50%\
    • delay line 100 m 
    • all fibers are PM type (polarity maintained)
    • Photodetector is "lab buddy", very fast diode.
    • Note: différance from schematic (we did not use a low pass filter)

 

 

ELOG V3.1.4-395e101