Message ID: 45     Entry time: Thu Dec 19 16:23:39 2019
 Author: Ronic Chiche 
 Status: Fixed 
 Type: info 
 Category: lasers and optics 
 Location: Optical room 
 Title: Cavity beam axis moving with power 

how the cavity beam axis is moving during a lock when the cavity is hot ?
could it explain a part of the Transmission / Coupling signal decay ?

we placed 2 Basler camera, one at (30+Z) cm and the other at (85+Z) cm (Z is about 15cm) from the M3 mirror, we recorded the video during a lock and we analyzed the centroid X and Y displacement at 2A and 3A.
frames acquisition speed is a quite slow ~ 100ms => we need to acquire the frames faster !

with these data, the displacement is no more than some pixels, which means << 100µm ... it should be completely negligeable for photdiode thorlabs DET100 with ~10mm of diameter.

the last picture displays typical locking curves (before and after lock) :
- transmission : yellow
- coupling : orange
- PZT correction : blue

Attachment 1: Basler_30cm_2A_v1.png  25 kB  | Show | Hide all | Show all
Attachment 2: Basler_30cm_2A_v2.png  28 kB  | Hide | Hide all | Show all
Basler_30cm_2A_v2.png
Attachment 3: Basler_30cm_3A_v1.png  28 kB  | Hide | Hide all | Show all
Basler_30cm_3A_v1.png
Attachment 4: Basler_30cm_3A_v2.png  26 kB  | Hide | Hide all | Show all
Basler_30cm_3A_v2.png
Attachment 5: Basler_85cm_2A_v1.png  27 kB  | Show | Hide all | Show all
Attachment 6: Basler_85cm_2A_v2.png  32 kB  | Show | Hide all | Show all
Attachment 7: Basler_85cm_3A_v1.png  30 kB  | Hide | Hide all | Show all
Basler_85cm_3A_v1.png
Attachment 8: Basler_85cm_3A_v2.png  32 kB  | Hide | Hide all | Show all
Basler_85cm_3A_v2.png
Attachment 9: Screenshot_2019-12-19_4_151242.png  45 kB  | Hide | Hide all | Show all
Screenshot_2019-12-19_4_151242.png
ELOG V3.1.4-395e101