HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports.  Not logged in ELOG logo
Entry   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics Capture_d'écran_2026-11-04_160149.pngCapture_d'écran_2026-11-04_160210.png
    Reply   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics IMG_20261104_164243(1).jpg
       Reply   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics IMG_20261106_160514(1).jpg
          Reply   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics RefCurve_2026-11-09_0_143449.pngRefCurve_2026-11-09_1_154254.png
             Reply   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics RefCurve_2026-11-09_2_161811.png
                Reply   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics untitled.png
                   Reply   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics RefCurve_2026-11-10_0_145752.png
                      Reply   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics 6x
                         Reply   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics RefCurve_2026-11-10_7_163813.pngRefCurve_2026-11-10_8_164036.png
                            Reply   Installation of a new cavity: the ABOX!!, posted by Alice Renaux at Optical room about lasers and optics untitled.png
Message ID: 266     Entry time: Fri Nov 6 16:30:36 2026     In reply to: 265     Reply to this: 267
 Author: Alice Renaux 
 Status: Under Process 
 Type: report 
 Category: lasers and optics 
 Location: Optical room 
 Title: Installation of a new cavity: the ABOX!! 

We estimated the FSR and the finesse:

- FSR: 161.037MHz

- Linewidth: 0.770ms x 20kHz/s = 15400kHz

- Finesse: FSR/linewidth = 10500

Far from the expected finesse. We will clean the mirrors and the vacuum chamber again.

Alice Renaux wrote:

ADD: We added the LaseLock to scan the NKT laser wavelength and select the transverse mode that resonates in the cavity before taking the attached picture.

**********

We installed a camera (Gentec-EO Beamage-4M) after M2 and a photodiode after M4 and at the reflection on M1 (ThorLabs DET100A2). We observed resonance peaks in transmission along with a drop in the reflected signal, which indicates some coupling (15-20% roughly ?) between the laser transverse mode and the cavity fundamental transverse eigenmode (see attachement).

We installed an EOM (Photline NIR-MPX-LM-0.1 S/N 4699-12). The attenuation coefficient is 3dB (drop of 50% in the laser power after the EOM).

We installed a PDH box to generate an error signal (N°2 ThomX).

 

Alice Renaux wrote:

We installed a 160MHz four-mirror OEC with two planar mirrors and two spherical mirrors. The radii of curvature of the spherical mirrrors are 0.5m.

The design distances are the following:

- Cavity length: 1.87m

- M1-M2 (planar mirrors): 411.62mm

- M3-M4 (spherical mirrors): 520.00mm

- axis: 70mm

- M1-M3 and M2-M4 (diagonals) : 471.04mm

- incidence angle : 149.16mrad

 

The design implementation in the vacuum chamber is the following:

- wall-M1 : 253.82mm

- wall-M3 : 199.63mm

 

The mirrors we used are:

- M1: C23017/1. Substrate: Fused silica. Absorption: <0.6ppm. Scattering: 2.5-3.5ppm. Transmission: 115ppm. Total losses: 117.5-119.1ppm. Radius of curvature: planar.

- M2 : C23018/4. Substrate: ULE. Absorption: <0.6ppm. Scattering: 3-7ppm. Transmission: 3ppm. Total losses: 6-10.6ppm. Radius of curvature: planar.

- M3 : C23018/11. Substrate: ULE. Absorption: <0.6ppm. Scattering: 3.8-4.5ppm. Transmission: 3ppm. Total losses: 6.8-8.1ppm. Radius of curvature: 0.5m.

- M4 : C23018/12: Substrate: ULE. Absorption: <0.6ppm. Scattering: 2.8-4ppm. Transmission: 3ppm. Total losses: 5.8-7.6ppm. Radius of curvature: 0.5m.

Calculating R_i=1-(A_i+S_i+T_i) with i=[1, 4] and R=√(R_1*R_2*R_3*R_4), the expected finesse is: F=π√R/(1−R)=43211-46164.

 

The laser source is : NKT Photonics Koheras Adjustik + collimator. The laser beam diameter at 1/e^2 after M2 is : 1.62mm in the x direction and 1.58mm in the y direction (see attachements).

 

 