HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports.  Not logged in ELOG logo
Entry   Neolase amplifier installation, posted by Ronic Chiche at Optical room about mechanics | lasers and optics | cabling 
    Reply   Neolase amplifier installation, posted by Ronic Chiche at Optical room about mechanics | lasers and optics | cabling 
       Reply   Neolase amplifier installation, posted by Ronic Chiche at Optical room about mechanics | lasers and optics | cabling 
          Reply   Neolase amplifier installation, posted by Ronic Chiche at Optical room about mechanics | lasers and optics | cabling 
             Reply   Neolase amplifier installation, posted by Ronic Chiche at Optical room about mechanics | lasers and optics | cabling 
                Reply   Neolase amplifier installation, posted by Ronic Chiche at Optical room about mechanics | lasers and optics | cabling 
                   Reply   Neolase amplifier installation, posted by Ronic Chiche at Optical room about mechanics | lasers and optics | cabling 
Message ID: 249     Entry time: Thu Aug 6 14:49:32 2026     In reply to: 248     Reply to this: 250
 Author: Ronic Chiche 
 Status: Under Process 
 Type: info 
 Category: mechanics | lasers and optics | cabling 
 Location: Optical room 
 Title: Neolase amplifier installation 

I drained the chiller to install again the short pipe (< 1m) in closed loop.

then, I filled again the chiller and did the compression process to obtain 1bar.
in that case, the flow measured by the chiller is correct about 5.8Lpm.

PS : the correct length for the previous calculation is 6m (2x 3m) and not 3m...
in that case, the flow should be ~ 15Lpm (and not 22 Lpm)

)Ronic Chiche wrote:

pipes are maybe too long ?

we can use the Darcy-Weisbach + Blasius laws (non linear laws).
I asked to ChatGPT, Claude, Mistral and MathGPT to solve it with these parameters :

L ~ 3m
D = 1 cm
water at 20°C
dP = 0.7 bar

they all found ~ 22-23L/min

so, the tube length should not be the problem... to be investiguated.

 

Ronic Chiche wrote:

now, I connected the chiller circuit to the amplifier, thanks to the adapters.

chiller IN => blue pipe => laser OUT
chiller OUT => black pipe => laser IN

after several filling and compression procedure, the flow is only 1.7Lpm and the pressure is 0.8 bar.
the temperature has been set to 20°C according to the manual operating range 18-23°C. 

the flow is too low !
according to the manual, the flow should be 3-5Lpm !

a very small leakage appears if the mechanical constrains on the pipe is too high.
so, I separated the IN and OUT pipes to release this constrain and fixed the pipes to keep the position.

 

Ronic Chiche wrote:

we have all the adaptor plugs for connecting the chiller.
so, I tested first the chiller connected in closed loop (on itself).

there is no manual, so procedure is not completely clear...

DO NOT PLUG ANYTHING BEFORE THE CHILLER ASK FOR IT
ONE MUST FOLLOW THE PROCEDURE ON THE CHILLER DISPLAY

1) if the chiller circuit is full of coolant liquid,
one must start by a System Startup -> Draining
(a trash tank and the hose will be need for emptying the circuit)

2) once the chiller circuit is empty
one must continue by System Startup -> Filling
(a clean demineralized water tank and the hose will be need for filling the circuit)

=> OK !
the chiller is able the fill and drain its circuit.
I saw the cooling water circulating in the closed loop.

 

 

 

 