HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports.  Not logged in ELOG logo
Entry   CELIA 100W laser amplifier repair , posted by Ronic Chiche at Optical room about lasers and optics 
    Reply   CELIA 100W laser amplifier repair , posted by Ronic Chiche at Optical room about lasers and optics IMG20241002144351.jpgIMG20241002144936.jpg20241002_161657.jpgClipboard_10-02-2024_01.jpg
       Reply   CELIA 100W laser amplifier repair , posted by Ronic Chiche at Optical room about lasers and optics Sans_titre.png
Message ID: 225     Entry time: Wed Oct 2 18:12:54 2024     In reply to: 224     Reply to this: 230
 Author: Ronic Chiche 
 Status: Under Process 
 Type: issue 
 Category: lasers and optics 
 Location: Optical room 
 Title: CELIA 100W laser amplifier repair  

this afternoon, we saw some electric cables badly connected to their power supply.
we fixed it by soldering them together and screwing the result to the power supply.
(see 1st image)

we lift the plate of the 1st stage and we check for optical leakage in the fibers (see 2nd image + picture of the top part of the cassette).
(Aurélien took several images)

without the 1st stage amplification, we saw some lealage only in the bottom part of the "optical cassette".
light was scattered mostly from one side (2 spots) and we saw also very weak scattering in the other directions.

with the 1st stage amplification, we clearly saw the losses from the bent fibers inside the top part of the cassette => it's a good sign.
but after the 5%-10% coupler (the one used for the diagnostic of the power to allow the use of the 2nd stage), we don't see any losses, which means there is no light in this part !
the fiber break could be in between...
Aurélien should send the images to Jérome to get a diagnostic.
the old schematic is attached but it has been modified in the Loic Thesis (p. 165)

we identified the black optics components as 2 isolators (AFW-PISO-30-1W-FB) and 1 circulator (AFW-CIR-PM-30) from AFW technologies.


 

Ronic Chiche wrote:

this morning with Alice, we sent the Menhir 160MHz injected in a fiber (with 6mW at the end of a long fiber) into the laser amplifier, to look for leakage or damage in the first stages of the amplifier (the amplifier is totally off).

first of all, we checked for light scattering around the laser crate with a sensitive optical card => nothing

and then, we checked for light scattering inside the laser crate with an optical viewer => we just saw 1 or 2 small spots located at the end of an optical element at the 2nd stage level.
but it's difficult to understand the optical path and know the different elements with the 1st stage still in place.

we think it is mandatory to open the top of the crate and lift the 1st stage to have a better look inside the optical parts which are at the 2nd and 3rd stage levels:
we could remove the front side of the crate without any damage to any fibers in the crate.

we just saw 1 fiber, glued to an optical element on the 2nd stage, and going to the 3rd stage.
the 1st stage is just an electronics parts stage which seems easy to be removed.

... to be discussed...

 

 

Attachment 1: IMG20241002144351.jpg  223 kB  | Hide | Hide all | Show all
IMG20241002144351.jpg
Attachment 2: IMG20241002144936.jpg  300 kB  | Hide | Hide all | Show all
IMG20241002144936.jpg
Attachment 3: 20241002_161657.jpg  316 kB  | Hide | Hide all | Show all
20241002_161657.jpg
Attachment 4: Clipboard_10-02-2024_01.jpg  168 kB  | Show | Hide all | Show all
ELOG V3.1.4-395e101