Entry   Mounts installation and cavity alignment, posted by Xinyi Lu at Optical room about lasers and optics 
    Reply   Mounts installation and cavity alignment, posted by Xinyi Lu at Optical room about lasers and optics CCD.jpgoscillograph.jpg
       Reply   Mounts installation and cavity alignment, posted by Xinyi Lu at Optical room about lasers and optics resonances.jpgcavity_mode.jpgcavity_mode_diameter.jpg
Message ID: 177     Entry time: Thu Dec 14 17:52:53 2023     In reply to: 176     Reply to this: 178
 Author: Xinyi Lu 
 Status: Fixed 
 Type: report 
 Category: lasers and optics 
 Location: Optical room 
 Title: Mounts installation and cavity alignment 

Over the past few days, Viktor, Ronic and I have continued to align the cavity. We installed 4 mirrors and monitored the transmitted laser with a CCD and photodiode. By adjusting the motors of the cavity mirror stages and the injector mirrors, we obtained resonances and less symmetric TEM20-like patterns. Possible reasons for this are a mismatch between the beam sizes of the laser and the cavity mode, and the mounts are installed in rough positions.
Tomorrow, we plan to use the CW laser to realign the optical cavity and position the mounts more precisely.

Xinyi Lu wrote:

Over the last two days, Viktor, Ronic and I have started to install the mirror mounts and try to align the cavity.
- We used the Menhir laser @ 160MHz for alignment.
- To make it easier to operate, we removed some lenses and waveplates, and kept only a few necessary reflective mirrors.
- We measured the distance with rulers and placed the mounts in designed positions.
- We installed Iris on the mirror mounts, used a CCD camera to determine if the beam was in the center, and optimized the two reflective mirrors outside the cavity.
- There were some problems with the controller of the injection mirrors (Newport™) in front of the cavity, and Ronic has fixed them successfully.
- Next week, we will continue to align the cavity, measure the cavity mode, and design the telescope. We will install the old SBX mirrors for alignment first, and then replace them with the final good mirrors.

 