Entry   Power storage In Cavity, posted by Manar Amer at Optical room about lasers and optics 
    Reply   Power storage In Cavity, posted by Manar Amer at Optical room about lasers and optics Transmission_after_S2.jpg
       Reply   Power storage In Cavity, posted by Manar Amer at Optical room about lasers and optics 
          Reply   Power storage In Cavity, posted by Manar Amer at Optical room about lasers and optics 00mode_higer.jpg00mode_Removed_higer.jpg
             Reply   liPower storage In Cavity, posted by Manar Amer at Optical room about lasers and optics Screenshot_2022-09-13_0_184445.png
                Reply   Power storage In Cavity, posted by Manar Amer at Optical room about lasers and optics High_Finesse_sign_on_transmission_.jpgScreenshot_2022-09-14_0_161520.png
                   Reply   Power storage In Cavity, posted by Manar Amer at Optical room about lasers and optics Screenshot_2022-09-14_0_161520.png
Message ID: 159     Entry time: Mon Sep 12 17:00:35 2022     In reply to: 158     Reply to this: 163
 Author: Manar Amer 
 Status: Fixed 
 Type: report 
 Category: lasers and optics 
 Location: Optical room 
 Title: Power storage In Cavity 

at max stage we reached at 2.5 A on the diodes we have

8000 gain , coupling ~ 60% , with power inside the cavity 50 kW

 

The higer order mode that was observed today,

it was also observed on Friday  along the other higher order mode recorded on the logbook.

after cutting it with the D-shaped mirror we recorded the shape (image attached)

Manar Amer wrote:

Power increased in the amplifier

reflective filters added at the transmission point to be able to have signal on the Beam Profiler, Transmission diode and power meter

Amplifier output current (A)

diode 1

Amplifier output current (A)

diode 2

Amplifier output current (A)

diode 3

Amplifier output current (A)

diode 4

Power Output (W) Transmitted power (mW) note
0.5 0.5 0.5 0.5     no change in the beam shape
1 1 1 1      
1.5 1 1 1     Appearance of higher order mode (vertical), suppressed using D-shaped mirror (vertical only) by (-23000), image recorded.
1.5 1.5 1.5 1.5     no change
2 1.8

1.5

1.5   52 mA 52 mA after CEP adjustment , power measured after 2 reflective fillers NDUV30A and NDUV20
2 2 2 2   70 mA  
2.5 2.5 2 2   95  
2.5 2.5 2.5 2.5    

sudden drop in power, cavity lock lost, lock system looks ok (maybe something in the cavity)

mode shape is the same

coupling is large, transmission is very low

change the CEP, mo effect in changing it (we have no gain in the cavity, No Finesse)  could be something happened to the mirrors !!!!!!

 

 

Manar Amer wrote:

At transmission power of 42 mW, coupling 60%, we see fundamental mode with a degeneracy

The D-shape motors were moved to a position 2 000 000 steps (in theory they should be the max position)

but no change appeared on the 00 mode or the higher order mode.

We will break the vacuum and check the position of the D-shaped mirrors.

Manar Amer wrote:

We start of power output from amplifier of 1W with only the first and second stage on then we start with the third stage power increase,

Note: when changing the current on the third stage 4 diodes, better to do it step by step for each one with a step of ~ 0.5 A

  •  

Amplifier output current (A)

for four diodes

Power Output (W)

Transmitted power (MW)

 

0 , 0 , 0 ,  0 ~ 1 2.3
1 , 1 , 1 , 1 / 11.0
1.5 , 1.3, 1 , 1 / 35
1.5 , 1.5 , 1 , 1 / 42.16
   

 

  • we placed at the transmission point a splitter the transmitted power before was 35 mW after placing it we had 26 mW
    • 97 % transmission , 3% reflected which goes to the beam profiler

 

  • We see degeneracy modes at the last step of 42.16 mW , so we started to test the D-shaped mirrors
    • the full range of the motors is 50.8 mm
    • each step =  30 nm , 30 000 steps = 1 mm

 

 

 

Attachment 1: 00mode_higer.jpg  270 kB  | Hide | Hide all | Show all
00mode_higer.jpg
Attachment 2: 00mode_Removed_higer.jpg  246 kB  | Show | Hide all | Show all