HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports.  Not logged in ELOG logo
Entry   Locking Amplifier with Cavity , posted by Manar Amer at Optical room about lasers and optics 20220907_00mode.jpg20220907_00mode_diameter.jpgScreenshot_2022-09-07_0_174654.pngScreenshot_2022-09-07_2_182520.png
    Reply   Locking Amplifier with Cavity , posted by Manar Amer at Optical room about lasers and optics 
       Reply   Locking Amplifier with Cavity , posted by Manar Amer at Optical room about lasers and optics Screenshot_2022-09-08_0_133245.png
          Reply   Locking Amplifier with Cavity , posted by Manar Amer at Optical room about lasers and optics Screenshot_2022-09-08_2_144338.png
             Reply   Locking Amplifier with Cavity , posted by Manar Amer at Optical room about lasers and optics 
             Reply   Locking Amplifier with Cavity , posted by Manar Amer at Optical room about lasers and optics 
                Reply   Locking Amplifier with Cavity , posted by Manar Amer at Optical room about lasers and optics Screenshot_2022-09-09_0_120737.png
                   Reply   Locking Amplifier with Cavity , posted by Manar Amer at Optical room about lasers and optics 
Message ID: 153     Entry time: Fri Sep 9 12:49:44 2022     In reply to: 152
 Author: Manar Amer 
 Status: Fixed 
 Type: report 
 Category: lasers and optics 
 Location: Optical room 
 Title: Locking Amplifier with Cavity  

Alignment was improved to ~ 1.2 V on the photodiode.

Only the second stage was on with ~ 1 W output from amplifier

we measured the transmission from S2 to be 2.3 mW

Manar Amer wrote:

image of the lock with Coupling of ~ 60%

 

Manar Amer wrote:

We locked the cavity, and it is stable using the Transmission,

the high frequency that we thought could have a reason for instability, is due to the high power on the photo diode of the PDH box which can cause non-linearity effects in the signal.

We also closed the fan of the third stage and there was no significant change on the error signal and the piezo+AOM compensation signal.  

        

Manar Amer wrote:

We managed to lock the cavity by adding the AOM , but the lock is still difficult to stabilize.

There is some high frequency compensated by the AOM at ~ 170 kHz (yet to understand from where it comes from)

Manar Amer wrote:

We played on the CEP using the USB command "id0=xxxxx" of the Menhir.

we put id0=48650 and we improved A LOT the transmission and coupling (~ 60%)

here is an image of the first attempts to lock... but the locking is quite difficult.
yellow : transmission
green : reflection
amplifier is still with 2nd stage only (Pout=~1W)

we are adding an AOM in the loop...

Manar Amer wrote:

Today, we played on the CEP but using the I-tune input of the Menhir Laser (+/- 5V maximal range).

unfortunately, one only saw a very weak improvement of the transmission by 10-20%... and the coupling improvement is almost zero.

the best improvement was for the maximal I-tune range (+5V) which maybe means that we could improve more the effect if were able to get a full range of 2pi for the CEP (instead of the present pi/2 range).

Manar Amer wrote:

Today we managed to observe the fundamental mode and stabilize the scan on it until we improved the alignment enough.

We see some coupling, but it is very week < 5% , We improved the alignment and the polarization, but there is still no explanation to why it is very low.

The mode shape is circular with radius = 0.89 mm at transmission point, ~ 40 cm from circular mirror.

One EOM was installed along the injection into the amplifier, we saw a drop in the power measured by the first stage monitor from ~ 8.2 mW to 7.2 mW

we improved the injection up to 7.6 mW, but it still fluctuates a lot. We need to be careful about it.

The error signal looks clean, but it is very week which is due to the weak coupling.

 

 

 

 

 

 

 

 

ELOG V3.1.4-395e101