Message ID: 8     Entry time: Tue Apr 27 13:57:15 2021     Reply to this: 9
 Author: Manar Amer 
 Status: Fixed 
 Type: report 
 Category: lasers and optics 
 Location: Other 
 Title: Alignment using Koheras  

Following the Helium Neon Alignment + change in the distance between the mirrors to be; M3-M4 = 90.5 mm , M1-M2 = 80.2 mm  -→ The alignment using the Koheras CW laser is done.

  • Additional components used:
    • for monitoring beam :  Photodiode (power of beam), Beam Profiler (shape, position, power , ... ) 
    • for Koheras frequency scan: function generator, Amplifier or use lase-lock (had some issues to be checked)
    • Telescope: made using 1 m focal length to match the beam shape of the cavity
  • Observed during:
    • The alignment is fairly similar to the previous one, placed two irises to preserve it.
    • Fundamental mode observed (beam profiler after M2) was circular
    • when the frequency scan was fine-tuned around the fundamental mode we could see the mode pulsing in the cavity, but there was a bit of instability.
    • when doing a very wide frequency scan (50 V ~ 1.5 GHz), multiple modes where showing inside the cavity

Photos attached show:

  • some resonating modes in the cavity 
  • Fundamental mode resonating in the cavity, with its properties (2D shape, 1D shape, position) ** The picture is taken after subtracting the background **
    • from 2D it is very circular
    • from 1D it is confirmed to be circular (715.00 um - 704.00 um)
    • From position, we have a reference to compare with tomorrow morning.
  • Diffraction that can be observed in the cavity (you can clearly see the edges of the mirrors in the photo)
  • The temperature curve is attached for the duration of the exp. (before starting Ronic switched something off and then put it on !!!!!! it seems to be for temp regulation )

** Notes for tomorrow morning :  first : switch on the laser and check if the beam 00 mode is still observed and check its position

this is to see the stability and the effect of the temperature.

Attachment 1: 20212704_Fundamental_mode_in_cavity_2D_-_Position.png  331 kB  | Hide | Hide all | Show all
20212704_Fundamental_mode_in_cavity_2D_-_Position.png
Attachment 2: 20212704_Fundamental_mode_in_cavity_1D_shape.png  231 kB  | Show | Hide all | Show all
Attachment 3: 20212704__Difraction_along_the_mirror_edges_in_cavity.png  193 kB  | Hide | Hide all | Show all
20212704__Difraction_along_the_mirror_edges_in_cavity.png
Attachment 4: 20212704__Modes_observed_resonating_in_the_cavity_0.png  236 kB  | Show | Hide all | Show all
Attachment 5: 20212704__Modes_observed_resonating_in_the_cavity_1.png  232 kB  | Hide | Hide all | Show all
20212704__Modes_observed_resonating_in_the_cavity_1.png
Attachment 6: 20212704_Temperature_Curve.jpg  78 kB  | Hide | Hide all | Show all
20212704_Temperature_Curve.jpg
ELOG V3.1.4-395e101