Entry   Cameras Calibration, posted by Kevin Dupraz at Optical room about lasers and optics | detectors and electronics NF_Refl_calib_hex2mm__2021-06-23__16-59-26.bmpNF_Trans_calib_hex2mm__2021-06-23__17-06-49.bmpNF_Inj_calib_DIODE__2021-06-24__10-56-00.bmpNF_Inj_calib_DIODE__2021-06-24__10-42-00.bmpNF_Trans__2021-06-25__16-38-56.bmp
    Reply   Cameras Calibration, posted by Kevin Dupraz at Optical room about lasers and optics | detectors and electronics grab2021-12-06_10-51-34_325_microscope.bmpgrab2021-12-06_10-52-11_108_microscope_calib_05mm_ruler.bmp
       Reply   Cameras Calibration, posted by Kevin Dupraz at Optical room about lasers and optics | detectors and electronics Plot_of_NF_Trans_calib_hex2mm__2021-06-23__17-06-49.pngPlot_of_NF_Refl_calib_hex2mm__2021-06-23__16-59-26.png
Message ID: 45     Entry time: Tue Dec 7 11:48:33 2021     In reply to: 44
 Author: Kevin Dupraz 
 Status: Under Process 
 Type: info 
 Category: lasers and optics | detectors and electronics 
 Location: Optical room 
 Title: Cameras Calibration 

After check a mistake has been found on the magnification. This seems to be the good calibration (feel free to cross check). The projection of the 2mm hex is attached

the calibration are :

NF_Refl: acA1920-40gm
    pixel size (real): 5.86um
    Magnification = 1.71
    pixel size (image): 10um
    image donne on input plan mirror M1 (accuracy about few mm)

NF_Trans: acA1920-40gm
    pixel size (real): 5.86um
    Magnification = 0.68
    pixel size (image): 4um
    image donne on output plan mirror M2 (accuracy about few mm)

The NF_inj was calibrated with the USB microscope (1.4um/pixel on microscope image). The 5th ring of the lens is about 1.89 mm in diameter.

NF_inj: acA1920-40gm
    pixel size (real): 5.86um
    Magnification = 1.54
    pixel size (image): 9um

Kevin Dupraz wrote:

 

Kevin Dupraz wrote:

After installing cameras the actual calibration are :

NF_Refl: acA1920-40gm
    pixel size (real): 5.86um
    Magnification = [0.53,0.56]
    pixel size (image): 3.22um
    image donne on input plan mirror M1 (accuracy about few mm)

NF_Trans: acA1920-40gm
    pixel size (real): 5.86um
    Magnification = 1.32
    pixel size (image): 7.73um
    image donne on output plan mirror M2 (accuracy about few mm)
 

need to adjust the NF transmission as the Magnification is greater than 1.

The NF_inj was calibrated with the USB microscope (1.4um/pixel on microscope image). The 5th ring of the lens is about 1.89 mm in diameter.

NF_inj: acA1920-40gm
    pixel size (real): 5.86um
    Magnification = 1.54
    pixel size (image): 9um

 

Attachment 1: Plot_of_NF_Trans_calib_hex2mm__2021-06-23__17-06-49.png  8 kB  | Hide | Hide all | Show all