Entry   Phase noise measurement of the Amplitude MIKAN, posted by Ronic Chiche at Optical room about lasers and optics | detectors and electronics 
    Reply   Phase noise measurement of the Amplitude MIKAN, posted by Ronic Chiche at Optical room about lasers and optics | detectors and electronics 
       Reply   Phase noise measurement of the Amplitude MIKAN, posted by Ronic Chiche at Optical room about lasers and optics | detectors and electronics MIKAN_optical_phase_noise_comparison_with_NKT_CW_oscillator_&_RIN_@54MHz.png
Message ID: 37     Entry time: Wed Oct 20 14:54:00 2021     In reply to: 2
 Author: Ronic Chiche 
 Status: Fixed 
 Type: info 
 Category: lasers and optics | detectors and electronics 
 Location: Optical room 
 Title: Phase noise measurement of the Amplitude MIKAN 

MIKAN phase noise and RIN measurements:

Ronic Chiche wrote:

Coupling into a 50-50% fiber coupler using the Thorlabs XYZ table NanoMax TS.
We reached 117mW after the 50% arm which means we coupled about 230mW (the coupling better than 50%).

Ronic Chiche wrote:

Installation of the MIKAN.
powermeter (with OD1) just after the oscillator shows 440mW for 4A of the MIKAN pump current.


Installation of a periscope as the oscillator beam position is very close to the table... not easy to put devices at this height (be aware to use proper wavelength range mirrors: Thorlabs BB1-E03)
Installation of a HALF waveplate to align beam polarisation on the isolator axis
Installation of a High power isolator (the one of ThomX): Newport ISO-FRDY-05-1030-N
powermeter (with OD1) just after the isolator shows 427mW

 

 

Attachment 1: MIKAN_optical_phase_noise_comparison_with_NKT_CW_oscillator_&_RIN_@54MHz.png  50 kB  | Show | Show all