  Status of commissioning, report also here plots are reports.  Not logged in ELOG logo
Message ID: 34     Entry time: Fri Oct 8 16:02:19 2021
 Author: Blanc 
 Status: Fixed 
 Type: info 
 Category: lasers and optics | detectors and electronics 
 Location: Optical room 
 Title: Setup and measurements 

To summerize the results of this week :

We made the setup in attachement 5 and align it to have the good mode in the cavity (TEM00).

With the good alignment, we found the lock of the GHz, in attachement 1. We used the motors in the cavity to position very precisely the mirrors to find this lock.

In attachement 2, we can see the error signal obtain thanks to the PDH, which is very clean.

But after that, we saw some modulations at 30 kHz frequency which represent the resonance frequency of the piezo in the GHz (attachement 3). To compensate this, we placed a fibered AOM to cut the high frequencies.

Finally, we observed a big difference in the transmission noise if the motor loop is open or closed (attachement 4). To have the best noise, we have to stop the closed loop and stay in open loop, after adjusting the mirrors with the cosed loop.

We have also measured the finesse (2000) and the coupling (20%), detailed in the previous post.

 

Attachment 1: acquisition_lock_GHz.png  36 kB  | Show | Hide all | Show all
Attachment 2: Signal_d'erreur_PDH_sur_GHz.png  34 kB  | Hide | Hide all | Show all
Signal_d'erreur_PDH_sur_GHz.png
Attachment 3: Résonances_dûes_au_pic_à_30kHz_dans_le_phase_noise.png  40 kB  | Show | Hide all | Show all
Attachment 4: Close_loop_vs_open_loop_GHz.png  37 kB  | Hide | Hide all | Show all
Close_loop_vs_open_loop_GHz.png
Attachment 5: minicav_setup_Lock_GHz.PNG  119 kB  | Show | Hide all | Show all